Chapter 4 –
Functions of the instrument
UM-23100B-U User
manual
ATEQ F CLASS
Page 52/90
e) "Fill" cycle end
If the part is OK, the good part relay is activated at the end of the test time and remains
so till the start of the next cycle.
At the end of the dump time, the end of cycle relay is activated.
If the part is bad, as soon as the test time is finished the relay becomes and remains
activated.
The instrument
waits for a reset from the operator or the PLC to start the dump time and
send the end of cycle signal.
PRESSURE
START
Reset
A
P-F
P-D
F
S
T
D
t
t
t
t
t
t
t
program
number
automatic
Connector A
t
Fail part
B time
B time
End of cycle
T
est
c
ycl
e
O
rder
s acti
ons
t mini
t mini
t mini
Active
Inactive
Active
Inactive
#
automatic
Connector B