41 / 74
FG_F6_MODBUS_EN_01 / 2020-06-17
Identifier N°
D.A. address
Meaning
Value
Dec
Hexa
Read
Write
53
0035
2035
6035
“Press. UNIT”
Pressure unit.
Refer to Unit table.
60
003C
203C
603C
“Test FAIL”
Natural reject value of the test part
0 > 9999
61
003D
203D
603D
“TestREWORK”
Natural reject level of the test part in
recovery
0 > 9999
62
003E
203E
603E
“Ref. FAIL”
Natural reject level of the reference part
0 > 9999
63
003F
203F
603F
“Ref.REWORK”
Natural reject value of the ref. part in
recovery
0 > 9999
66
0042
2042
6042
“Set FILL”
Fill instruction value:
- 9999 > 9999
67
0043
2043
6043
“Set PreFILL“
Pre-fill instruction value:
- 9999 > 9999
68
0044
2044
6044
“SEALED PART”
Choice of the sealed component
Standard
Large Leak
0000
1000
72
0048
2048
6048
“Drift Unit”
Calibration drifts percent.
0 > 100%
80
0050
2050
6050
“Diff A
-
Z”
Differential auto reset time.
0 > 650 seconds
102
0066
2066
6066
“BLOW MODE”
Type of permanent blowing
Regulator 2
Regulator 1
0000
1000
103
0067
2067
6067
“FILL MODE”
Type of fill.
Standard
Instruction
Ballistic
Ramp
Adjust
Auto-Fill
Ramp 2
EASY
EASY Auto
0000
1000
2000
3000
4000
5000
6000
7000
8000
104
0068
2068
6068
“PreFILL”
Type of pre-fill.
Standard
Instruction
Ballistic
Ramp
EASY
EASY Auto
0000
1000
2000
3000
4000
5000
106
006A
206A
606A
“CheckTime”
Commutation time of the equalization valve
calibration check.
0 > 650 seconds
107
006B
206B
606B
“% Drift”
ATR absorption tolerance.
0 > 100%
108
006C
206C
606C
“Start”
Start value of the transient (ATR).
- 9999 > 9999
110
006E
206E
606E
“EXT. DUMP”
Type of external dump.
Normally
close
Normally
open
0000
1000
111
006F
206F
606F
“Ref. VOL.“
Reference volume.
0 > 9999