FG_F6_MODBUS_EN_01 / 2020-06-17
42 / 74
Identifier N°
D.A. address
Meaning
Value
Dec
Hexa
Read
Write
112
0070
2070
6070
‘IN7:”
Function attributed to the entry of the
special cycles (input 7)
Refer to the
“
Configurable input
values
” table at the
end of this chapter
117
0075
2075
6075
“Set Blow”
Permanent blowing pressure instruction.
- 9999 > 9999
118
0076
2076
6076
“REJECT CALC.”
Original unit for the calculation of rejects in
cm3/ min (Pa, Pa/s...).
Refer to Unit table.
119
0077
2077
6077
“Min Level” (Sealed Diff menu)
Minimum sealed component measurement
pressure.
0 > 9999
120
0078
2078
6078
“Max Level” (Sealed Diff menu)
Maximum sealed component measurement
pressure.
0 > 9999
121
0079
2079
6079
“FILL TIME”(Sealed Diff menu) Fill time of the
internal volume.
0 > 650 seconds
122
007A
207A
607A
“TRANSFER”(Sealed Diff menu) Sealed
component transfer time.
0 > 650 seconds
123
007B
207B
607B
“LANGUAGE”
Choice of the language.
Default
language
2
nd
predefined
language
0000
1000
124
007C
207C
607C
“Max Value”
Reject in calibration check.
0 > 9999
125
007D
207D
607D
“% Drift”
Percentage of the calibration check.
0 > 100%
126
007E
207E
607E
“Max PreFILL”
Maximum pressure value in pre-fill.
- 9999 > 9999
127
007F
207F
607F
“LeakUnit”
Reject unit.
Refer to Unit table
128
0080
2080
6080
“Leak Rate”
Instruction value during a calibration.
0 > 9999
135
0087
2087
6087
“% of T FAIL”
Reject level percent of the auto parameter
function
0 > 100%
138
008A
208A
608A
“FILL REG”
Regulator number selection for the fill.
Regulator 1
Regulator 2
0000
1000
139
008B
208B
608B
“PRE FILL REG”
Regulator number selection for the pre-fill.
Regulator 1
Regulator 2
0000
1000
140
008C
208C
608C
“CORRECTION” (TEMP.CORR. 1 menu)
Percentage concerning the temperature
compensation.
0 > 100%
141
008D
208D
608D
“TEST TIME”(TEMP.CORR. 1 menu)
Test time for the temperature
compensation.
0 > 650 seconds
142
008E
208E
608E
“Max FILL”
Max pressure in indirect test (piezo 2).
- 9999 > 9999
143
008F
208F
608F
“Min FILL”
Min pressure in indirect test (piezo 2).
0 > 9999
144
0090
2090
6090
“OUTPUTS CONFIG.”
Setup of the outputs (standard or compact).
Standard
Compact
0000
1000