43 / 74
FG_F6_MODBUS_EN_01 / 2020-06-17
Identifier N°
D.A. address
Meaning
Value
Dec
Hexa
Read
Write
148
0094
2094
6094
“FILTER”
Filtering.
0 > 650 seconds
149
0095
2095
6095
“UNITS”
Unit type
SI
SAE
CUSTOM
0000
1000
2000
161
00A1
20A1
60A1
“Volume UNIT”
Volume unit.
Refer to Unit table.
164
00A4
20A4
60A4
“NEXT PROG.”
Number of the following program in
sequencing.
1 > 128
165
00A5
20A5
60A5
“N. OF CYCLES”(PIEZO AUTO AZ menu)
Number of cycles between two automatic
reset.
0 > 9999
166
00A6
20A6
60A6
“N. OF MINUTES”(PIEZO AUTO AZ menu)
Time between two automatic reset.
0 > 999 minutes
175
00AF
20AF
60AF
“REGUL. CTRL.”
Regulator check during its learning.
Automatic
Ext
0000
1000
203
00CB
20CB
60CB
“ELEC. REG.”
Activation or not of the built in electronics
regulators.
None
Reg 1
Reg 2
ALL Reg
0000
1000
2000
3000
232
00E8
20E8
60E8
“ATR DRIFT”
Drift transient (ATR).
0 > 100%
233
00E9
20E9
60E9
“AZ SHORT”
Quick auto-zero time.
0 > 650 seconds
273
0111
2111
6111
“DUMP”
Dump time in calibration check mode
0 > 650 seconds
291
0123
2123
6123
“T.ATR2”
Stabilization time for the ATR 2 function
0 > 650 seconds
295
0127
2127
6127
“DUMP LEVEL”
Minimum dump pressure level to reach
- 9999 > 9999
297
0129
2129
6129
“MAX BLOW”
Blowing maximum pressure level
- 9999 > 9999
298
012A
212A
612A
“MIN BLOW”
Blowing minimum pressure level
- 9999 > 9999
315
013B
213B
613B
“Start FILL”
Start value of the fill instruction in burst test
mode
- 9999 > 9999
334
014E
214E
614E
“RAMP”
Rise time in burst test mode
0 > 650 seconds
335
014F
214F
614F
“T. LEVEL”
Step time in burst test mode
0 > 650 seconds
336
0150
2150
6150
”N. OF STEPS”
Step number in burst test mode
0 > 650 seconds
340
0154
2154
6154
“Transient”
ATR transient value.
- 9999 > 9999
349
015D
215D
615D
“FILL TIME” (Indirect menu)
Fill time in recovery test mode
0 > 650 seconds
353
0161
2161
6161
“Press. UNIT”
(configuration/pneumatic menu)
General pressure unit
Refer to Unit table.
354
0162
2162
6162
“LINE P. MIN”
Minimum line pressure level
- 9999 > 9999