FG_F6_MODBUS_EN_01 / 2020-06-17
44 / 74
Identifier N°
D.A. address
Meaning
Value
Dec
Hexa
Read
Write
355
0163
2163
6163
“FILL TIME”
(AUTO VOL menu)
Internal volume fill time in program selection
by volume function
0 > 650 seconds
356
0164
2164
6164
“TRANSFER” (AUTO VOL menu)
Transfer time in program selection by volume
function
0 > 650 seconds
357
0165
2165
6165
“DUMP TIME”
(AUTO VOL menu)
Dump time in program selection by volume
function
0 > 650 seconds
358
0166
2166
6166
“PRESSU. VOL” (AUTO VOL menu)
Internal volume in program selection by
volume function
0 > 9999
359
0167
2167
6167
“Ref. VOL.” (AUTO VOL menu)
Reference volume in program selection by
volume function
0 > 9999
360
0168
2168
6168
“INT REF VOL” (AUTO VOL menu)
Internal reference volume in program selection
by volume function
0 > 9999
361
0169
2169
6169
“INT TEST VOL” (AUTO VOL menu)
Internal test volume in program selection by
volume function
0 > 9999
362
016A
216A
616A
“VOL. STEP“ (AUTO VOL menu)
Volume slice in program selection by volume
function
0 > 9999
363
016B
216B
616B
“DUMP TIME” (Sealed Diff menu)
Dump time in sealed components
0 > 650 seconds
364
016C
216C
616C
“DISPLAY MODE“
Leak display management
xxxx
xxx.x
xx.xx
x.xxx
0000
1000
2000
3000
366
016E
216E
616E
“MODE” (EXT DUMP menu)
Dump mode
Continuou
s
Time
0000
1000
367
016F
216F
616F
“Program” (DUMP OFF menu)
Program number of the dump of function
0 > 128
368
0170
2170
6170
“Tolerance A”
Tolerance level A for ntest cycle
0 > 100%
369
0171
2171
6171
“Tolerance B”
Tolerance level B for ntest cycle
0 > 100%
370
0172
2172
6172
“OFFSET”(TEMP.CORR. 1 menu)
Temperature correction offset
- 9999 > 9999
371
0173
2173
6173
“NAME:”(Units menu)
CAL unit personalization
CHAR[5]
372
0174
2174
6174
“BYPASS”
Bypass valve mode selection
Pre-Fill +
Fill
Pre-Fill
Fill
0000
1000
2000
373
0175
2175
6175
“% Cut OFF”
Cut off function Percent
0 > 100%
374
0176
2176
6176
“ATF TIME”
Divisor time of ATF function
0 > 650 seconds
375
0177
2177
6177
‘IN
8
:”
Function attributed to the entry of the
special cycles (input 8)
Refer to the
“Configurable input
values” table at the
end of this chapter