45 / 74
FG_F6_MODBUS_EN_01 / 2020-06-17
Identifier N°
D.A. address
Meaning
Value
Dec
Hexa
Read
Write
376
0178
2178
6178
‘IN
9
:”
Function attributed to the entry of the
special cycles (input 9)
Refer to the
“Configurable input
values” table at the end
of this chapter
377
0179
2179
6179
“MEAS. START”
Waiting time for starting the
measurement in burst test
0 > 650 seconds
378
017A
217A
617A
“Time Adj”
Adjusting fill time (electronic regulator)
0 > 650 seconds
379
017B
217B
617B
“USB:”
USB mode (printer or supervision)
Supervision
Printer
Bar code
Auto
None
0000
1000
2000
3000
4000
380
017C
217C
617C
“Press. UNIT“(Indirect menu)
Pressure unite for recovery test
Refer to Unit table
405
0195
2195
6195
“TRANSF.TIME” (Sealed Diff menu) Sealed
Diff, Transfer time.
0 > 650 seconds
406
0196
2196
6196
“PRESS.CORR.” (Sealed Diff menu) Sealed
Diff, Pressure Correction.
- 9999 > 9999
407
0197
2197
6197
“LARGE LEAK” (Sealed Diff menu) Sealed
Diff, Large Leak Max.
0 > 9999
408
0198
2198
6198
“OFFSET” (Sealed Diff menu)
Sealed Diff, Offset.
- 9999 > 9999
409
0199
2199
6199
“FILL MODE”(Indirect menu)
Type of fill Reg 2.
EASY
EASY Auto
0000
1000
410
019A
219A
619A
“DUMP TIME” (Indirect menu)
Indirect Dump Time
0 > 650 seconds
455
01C7
21C7
61C7
“DROP PRESS.%”
Drop Press function Percent
0 > 100%
456
01C8
21C8
61C8
“ATM PRESS.“
Atmospheric Pressure
900 > 1100
457
01C9
21C9
61C9
“TEMP.”
Temperature
0 > 800
458
01CA
21CA
61CA
“DISP. OPT.”
Display Option in flow reject
None
Pa Display
Ambient Temp.
Object Temp.
Test check
ATR
Temp.
correction
Leak offset
learning
PATM
correction
0000
1000
2000
3000
4000
5000
6000
7000
8000
459
01CB
21CB
61CB
“N. OF CYCLES”
Number of learning cycle
2 > 9999
460
01CC
21CC
61CC
“INTER
-
CYCLE”
Time between 2 learning cycle
0 > 650 seconds
461
01CD
21CD
61CD
“MAX OFFSET”
Offset max for learning cycle
0 > 9999