61 / 74
FG_F6_MODBUS_EN_01 / 2020-06-17
Auto-zero on the ATEQ device
This functionality is only available in Standard access.
Master
Slave
—
Select the program on which you want to make the
auto zero
—
Make a
Write N*words
request of 1 word at the
address
0201h
and pass the value of the auto zero
special cycle (n°7).
On network:
01
10
02
01
00
01
02
07
00
86
71
01
Slave address
10
Function number (Write N*words)
02 01
Word address (Special cycle)
00 01
Number of words to write
02
Number of bytes to write
07 00
Word 1: write 0007h (spe. cycle n°7)
86 71
CRC
—
Answer to the request:
On network:
01
10
02
01
00
01
51
B1
01
Slave address
10
Function number (Write N*words)
02 01
Word address (Special cycle)
00 01
Number of written words
51 B1
CRC
—
Make a start request to launch the special cycle.