67 / 74
FG_F6_MODBUS_EN_01 / 2020-06-17
Reset FIFO results
This functionality is only available in Standard access.
This command resets the 8 last cycle’s results available in the FIFO.
Master
Slave
—
Make a
Write bit
request at the address
0002h
and force the Reset FIFO bit to 1
.
On network:
01
05
00
02
FF
00
2D
FA
01
Slave address
05
Function number (Write bit)
00 00
Bit address (Reset FIFO bit)
FF 00
Force bit to 1
2D FA
CRC
—
Answer to the request (exactly the same as the
request):
On network:
01
05
00
02
FF
00
2D
FA
01
Slave address
05
Function number (Write bit)
00 00
Bit address (Reset FIFO bit)
FF 00
Force bit to 1
2D FA
CRC