69 / 74
FG_F6_MODBUS_EN_01 / 2020-06-17
Words
D.A. address
Read
Meaning
Type
Bytes
Coeff
V2.xxx Only
25
2319
Pa
–
Pa/s Leak result low part word
Long
4
x1000
26
231A
Pa
–
Pa/s Leak result high part word
27 - 36
Unused
37
2325
Atmospheric pressure in hPa low part word
Long
4
x1000
38
2326
Atmospheric pressure in hPa high part word
39
2327
Temperature in °C low part word
Long
4
x1000
40
2328
Temperature in °C high part word
All the numerical values are treated with Long format with fixed comma (10
-3
). Thus, they must be
multiplied by 1000 to get the value in units (see examples in “Basic notions” section).