background image

 

www.atestgaz.pl

In case of improper operation of the system recheck all wiring and connections or contact
the system provider /equipment manufacturer. 

The gas monitoring system can be approved for operation only after passing results of all

aforementioned checks and verifications. 

7.5

Configuration of gas detectors

Reliable   operation   of   gas   detector   needs   preliminary   configuration   of   the   device   parameters.   The
configuration is performed with the use of two buttons 



(CONFIG) that are available after opening of the

device housing (see Figure 1). These buttons enable also displaying of the detector settings. 
Information about the detector settings is displayed on side wall of the device housing (see Figure 1).

7.5.1

Assigning addresses to gas detectors

Depress the arrow buttons  



  (see Figure  1)  and hold them about one second to enter the mode of

address assignment (the ‘Ad’ message appears on the device display). Release the buttons with the ‘Ad’

message active and the message starts blinking slowly – the desired address can now be set by means of
the arrow buttons 



. Finally, depress the both arrow buttons 



 simultaneously to confirm the address

and store it into the device (the setting continues blinking for about two seconds). Finally, the device quits
the procedure of address assignment. 
Should the selected address is not confirmed and stored, the old address remains intact and after 15

seconds the detector quits the mode of address assignment. 
The default address is 99. 

7.5.2

Operation modes of gas detectors

The two possible operation modes can be pre-set: 

the detector is operates as an independent device, 
the detector is coupled with the Teta MiniDet device (default setting). 

Depress the arrow buttons 



(see figure 1) and hold them about two seconds to enter the procedure that

enables selection of the operation mode (the ‘HE’ message appears on the device display). Release the

buttons with the ‘HE’ message active and the message starts blinking slowly – the desired operation mode
can now be set by means of the arrow buttons 



. There are two possible options where:

‘0’ means independent operation of the detector, 
‘1’ stands for collaboration with the Teta MiniDet device. 

Simultaneous depressing of the both buttons confirms the selection and the operation mode is stored in the

detector memory (the setting continues blinking for about two seconds). Finally, the device quits the
procedure of the operation mode selection.
Omission to confirm the selected operation mode leaves the previous selection intact in the device memory

after 15 seconds the detector quits the procedure of operation mode selection. 

When the gas detector is configured for independent operation but the Teta MiniDet device
is connected anyway, an error message appears (indicated by the LPG LED).

s. 14|20

User Manual: POD-046-ENG R01

Summary of Contents for Teta EcoN

Page 1: ...POD 046 ENG R01 User Manual Gas Detector Teta EcoWent Teta EcoN Product code PW 105 CO PW 111 NO2...

Page 2: ...Monitoring Detection and Reduction of gas hazards For more details please visit our website www atestgaz pl Atest Gaz A M Pachole sp j ul Spokojna 3 44 109 Gliwice Poland tel 48 32 238 87 94 fax 48 3...

Page 3: ...ces in not allowed Engineering of a gas detection system for any specific facilities to be safeguarded may need consideration of other requirements during the entire lifetime of the product Use of una...

Page 4: ...18 8 Technical specification 19 9 List of fast wearing parts 19 10 List of accessories 20 11 Marking codes of products 20 12 Appendices 20 List of Tables Table 1 Electric connections 9 Table 2 Indica...

Page 5: ...sholds Indication of detector faults 1 2 Operation principle Gas detectors are crucial components of any system for gas detection They are designed for periodical measurements of gas concentration in...

Page 6: ...indications preferably under daily basis periodical inspection and calibration see Section 7 7 1 1 2 2 Alarm 1 threshold 1 Alarm 2 threshold 2 It is the device status when only slight concentrations o...

Page 7: ...edure takes about two minutes and the detector starts normal operation 2 Safety Do never install gas detectors at locations exposed to direct effect of water moisture or direct sunlight When defects a...

Page 8: ...entional painting or paint splashing Before use of silicon or silicon based materials paints adhesives sealant etc make sure that the device is properly secured against unintentional coating 3 Device...

Page 9: ...Ds are located on the side wall of the device housing see Figures 1 and 2 Figure 2 User interface Table 2 summarizes how status of the gas detector is indicated by means of individual LEDs Detailed de...

Page 10: ...g it is just enough to approach a permanent magnet to a specific location on the housing this location is shown in Figure 1 Depending on the operation mode of the detector the following details are di...

Page 11: ...hat type If the original package or another protecting means e g corks is unavailable the conveyed equipment must be secured against shocks vibrations or moisture by means of adequate methods and mate...

Page 12: ...Figure 1 Drilling of mounting holes in walls or other permanent structures is easier with use of a drilling pattern attached to the device package 7 3 1 Electric network The electric network must be d...

Page 13: ...all devices carry out the following operations assign addresses to all detectors see details in section 7 5 1 select the operation mode s see details in section 7 5 2 configure the control unit see Us...

Page 14: ...lected address is not confirmed and stored the old address remains intact and after 15 seconds the detector quits the mode of address assignment The default address is 99 7 5 2 Operation modes of gas...

Page 15: ...e operation mode see Section 7 5 2 Gas detector is coupled with the Teta MiniDet device Defect of Teta MiniDet Gas Detector Improper setting for the operation mode see Section 7 5 2 Improper connectio...

Page 16: ...f the manufacturer to come and calibrate the detectors on site 7 7 2 Replacement of fast wearing parts Please refer to Table 6 for the recommended lifetime and replacement schedule of fast wearing par...

Page 17: ...out any delay When the detector under test is coupled with the Teta MiniDet detectors the test gas LPG must be also supplied to the second device Correct execution of the test needs application of a c...

Page 18: ...magnet to gas detector or initiation of alarm simulations 7 7 4 Maintenance Gas detectors should be cleaned only by wiping its housing with a soft damp cloth Cleaning agents containing solvents white...

Page 19: ...sed into the installation pipe diameter of the pipe 16 mm Glands multiband diameter of wire 3 5 12 mm Acceptable cable cores 0 2 2 5 mm2 wire cores 0 2 2 5 mm2 stranded cores Enclosure material ABS We...

Page 20: ...ized reference gas nitrogen dioxide NO2 with concentration of 10 ppm Table 7 List of accessories 11 Marking codes of products Product code Device name PW 105 CO Gas Detector Teta EcoWent PW 111 NO2 Ga...

Page 21: ...ng to electromagnetic compatibility EN 50270 2006 This declaration of conformity is issued under the sole responsibility of the manufacturer Purpose and scope of use product is intended for use in gas...

Page 22: ...atmosphere 4 3 1 2 Impact of the presence of a working gas or other reaction gas 4 3 1 3 Effect of oxygen concentration value 5 3 2 Influence of temperature 5 3 3 Influence of moisture 5 3 4 Influenc...

Page 23: ...reference electrode is used 5 It has a constant potential that is not dependent on the concentration of the measured gas Owing to electrode reactions electrons are generated whose orderly flow electri...

Page 24: ...ith the provisions in the User s Manual 3 1 Composition of the controlled atmosphere It is assumed that the standard composition of the atmosphere is a mixture of gases with the proportions indicated...

Page 25: ...signal does not change e g nitrogen at a carbon monoxide sensor A set of substances exists or may exist for every detector the ones that are inert to the sensor do not react chemically with the sensor...

Page 26: ...ed atmosphere is within the rated range of the detector operation it does not affect the reading However sudden abrupt changes of humidity in the measured atmosphere can cause momentary appearance of...

Page 27: ...rocess e g weather conditions safety potential hazard generated by the mixture criterion B These both criteria have an impact on the final cost of the calibration services and level of competence requ...

Page 28: ...of flammable components 60 LEL concentration of toxic components NDSCh2 oxygen concentration 25 vol tank 3 dm3 water capacity and p 70 atm specified liquid chemical compounds e g glycerol 1 2 propaned...

Page 29: ......

Page 30: ......

Page 31: ......

Page 32: ......

Reviews: