background image

www.atestgaz.pl

Ingredient

C [% vol]

C [ppm]

Nitrogen

78.084

780 840

Oxygen

20.946

209 460

Argon

0.934

9 340

Carbon dioxide

0.0360

360

Neon

0.00181

18.18

Helium

0.00052

5.24

Methane

0.00017

1.70

Krypton

0.00011

1.14

Hydrogen

0.00005

0.50

Xenon

0.000008

0.087

Table 2: Typical composition of atmospheric air

When gas concentrations in the atmosphere differ significantly from those indicated in the table  2, it is

necessary to conduct the analysis of the impact of such a situation on the detector operation.
This publication assumes that the detector is operated under a standard formulation.

3.1.1

Impact of the presence of a substance in a controlled atmosphere

The electrochemical sensor is relatively selective. This means that in addition to the "working gas," that is

the one which the detector responds to, these sensors also respond to a relatively small number of other
substances in comparison to other measurement techniques.

Because of the indication, a set of substances exists or can exist:

to which the sensor responds 

in plus

 giving a positive signal proportional to the concentration of a

substance (e.g. carbon monoxide on a carbon monoxide sensor, being its "working gas", hydrogen

on a carbon monoxide sensor),

to which the sensor responds 

in minus

 giving a negative signal proportional to the concentration of

a substance (for example, nitrogen dioxide for a sulphur dioxide sensor),

to which sensor does not respond (the output signal does not change – e.g., nitrogen at a carbon

monoxide sensor).

A set of substances exists or may exist for every detector:

the ones that are inert to the sensor – do not react chemically with the sensor detector element

(e.g. nitrogen on a carbon monoxide sensor),

the   ones   that   react   chemically   with   the   detector   element   of   the   sensor   and   do   not   cause

a supernormal degradation of its properties (e.g. sensor working gas),

the ones that affect the sensor elements and cause temporary or permanent degradation of its

characteristics or properties.

Due to the physio-chemical phenomena, for each sensor there may be a collection of substances which

interact with the working substance leaving no possibility of reaching the detector elements of the sensor
by the working gas (e.g. ammonia dissolves in water vapour). In the case of some substances (e.g. chlorine,

phosgene, hydrogen chloride, sulphur dioxide) the operating range of the sensor is low enough so that the
gas at these concentrations can be absorbed by water vapour from the atmosphere, or condensed on the

elements   of   the   sensor.   Thus,   it   will   not   be   the   same   one,   visible   for   the   detector,   as   long   as   its
concentration does not reach a sufficiently high value and does not saturate water vapour being on the way

to the sensor.

3.1.2

Impact of the presence of a working gas or other reaction gas

The reaction of gas from the detector element of the sensor causes its wear, therefore, detectors equipped

Atest-Gaz

 

A. M. Pachole sp. j.

ul. Spokojna 3, 44-109 Gliwice, Poland

VAT No.: PL 9691433231

tel.: +48 32 238 87 94

fax: +48 32 234 92 71

e-mail: contact@atestgaz.pl

p. 4/5

Appendix: PU-Z-032-ENG R02

Summary of Contents for Teta EcoN

Page 1: ...POD 046 ENG R01 User Manual Gas Detector Teta EcoWent Teta EcoN Product code PW 105 CO PW 111 NO2...

Page 2: ...Monitoring Detection and Reduction of gas hazards For more details please visit our website www atestgaz pl Atest Gaz A M Pachole sp j ul Spokojna 3 44 109 Gliwice Poland tel 48 32 238 87 94 fax 48 3...

Page 3: ...ces in not allowed Engineering of a gas detection system for any specific facilities to be safeguarded may need consideration of other requirements during the entire lifetime of the product Use of una...

Page 4: ...18 8 Technical specification 19 9 List of fast wearing parts 19 10 List of accessories 20 11 Marking codes of products 20 12 Appendices 20 List of Tables Table 1 Electric connections 9 Table 2 Indica...

Page 5: ...sholds Indication of detector faults 1 2 Operation principle Gas detectors are crucial components of any system for gas detection They are designed for periodical measurements of gas concentration in...

Page 6: ...indications preferably under daily basis periodical inspection and calibration see Section 7 7 1 1 2 2 Alarm 1 threshold 1 Alarm 2 threshold 2 It is the device status when only slight concentrations o...

Page 7: ...edure takes about two minutes and the detector starts normal operation 2 Safety Do never install gas detectors at locations exposed to direct effect of water moisture or direct sunlight When defects a...

Page 8: ...entional painting or paint splashing Before use of silicon or silicon based materials paints adhesives sealant etc make sure that the device is properly secured against unintentional coating 3 Device...

Page 9: ...Ds are located on the side wall of the device housing see Figures 1 and 2 Figure 2 User interface Table 2 summarizes how status of the gas detector is indicated by means of individual LEDs Detailed de...

Page 10: ...g it is just enough to approach a permanent magnet to a specific location on the housing this location is shown in Figure 1 Depending on the operation mode of the detector the following details are di...

Page 11: ...hat type If the original package or another protecting means e g corks is unavailable the conveyed equipment must be secured against shocks vibrations or moisture by means of adequate methods and mate...

Page 12: ...Figure 1 Drilling of mounting holes in walls or other permanent structures is easier with use of a drilling pattern attached to the device package 7 3 1 Electric network The electric network must be d...

Page 13: ...all devices carry out the following operations assign addresses to all detectors see details in section 7 5 1 select the operation mode s see details in section 7 5 2 configure the control unit see Us...

Page 14: ...lected address is not confirmed and stored the old address remains intact and after 15 seconds the detector quits the mode of address assignment The default address is 99 7 5 2 Operation modes of gas...

Page 15: ...e operation mode see Section 7 5 2 Gas detector is coupled with the Teta MiniDet device Defect of Teta MiniDet Gas Detector Improper setting for the operation mode see Section 7 5 2 Improper connectio...

Page 16: ...f the manufacturer to come and calibrate the detectors on site 7 7 2 Replacement of fast wearing parts Please refer to Table 6 for the recommended lifetime and replacement schedule of fast wearing par...

Page 17: ...out any delay When the detector under test is coupled with the Teta MiniDet detectors the test gas LPG must be also supplied to the second device Correct execution of the test needs application of a c...

Page 18: ...magnet to gas detector or initiation of alarm simulations 7 7 4 Maintenance Gas detectors should be cleaned only by wiping its housing with a soft damp cloth Cleaning agents containing solvents white...

Page 19: ...sed into the installation pipe diameter of the pipe 16 mm Glands multiband diameter of wire 3 5 12 mm Acceptable cable cores 0 2 2 5 mm2 wire cores 0 2 2 5 mm2 stranded cores Enclosure material ABS We...

Page 20: ...ized reference gas nitrogen dioxide NO2 with concentration of 10 ppm Table 7 List of accessories 11 Marking codes of products Product code Device name PW 105 CO Gas Detector Teta EcoWent PW 111 NO2 Ga...

Page 21: ...ng to electromagnetic compatibility EN 50270 2006 This declaration of conformity is issued under the sole responsibility of the manufacturer Purpose and scope of use product is intended for use in gas...

Page 22: ...atmosphere 4 3 1 2 Impact of the presence of a working gas or other reaction gas 4 3 1 3 Effect of oxygen concentration value 5 3 2 Influence of temperature 5 3 3 Influence of moisture 5 3 4 Influenc...

Page 23: ...reference electrode is used 5 It has a constant potential that is not dependent on the concentration of the measured gas Owing to electrode reactions electrons are generated whose orderly flow electri...

Page 24: ...ith the provisions in the User s Manual 3 1 Composition of the controlled atmosphere It is assumed that the standard composition of the atmosphere is a mixture of gases with the proportions indicated...

Page 25: ...signal does not change e g nitrogen at a carbon monoxide sensor A set of substances exists or may exist for every detector the ones that are inert to the sensor do not react chemically with the sensor...

Page 26: ...ed atmosphere is within the rated range of the detector operation it does not affect the reading However sudden abrupt changes of humidity in the measured atmosphere can cause momentary appearance of...

Page 27: ...rocess e g weather conditions safety potential hazard generated by the mixture criterion B These both criteria have an impact on the final cost of the calibration services and level of competence requ...

Page 28: ...of flammable components 60 LEL concentration of toxic components NDSCh2 oxygen concentration 25 vol tank 3 dm3 water capacity and p 70 atm specified liquid chemical compounds e g glycerol 1 2 propaned...

Page 29: ......

Page 30: ......

Page 31: ......

Page 32: ......

Reviews: