Q46H/65 Residual Chlorine Dioxide System
Part 6
– Configuration
36
O&M Manual
Rev-J (11/18)
6.22 Measure Menu [MEASURE]
The default menu for the system is the display-only menu MEASURE. This menu is a display-only
measurement menu, and has no changeable list items. When left alone, the instrument will
automatically return to this menu after approximately 30 minutes. While in the default menu, the
UP arrow allows the user to scroll through the secondary variables on the lower line of the
display. A brief description of the fields in the basic analyzer version is as follows:
ANALYZER MEAS SCREENS:
25.7°C
Temperature display can be displayed in °C or °F, depending on user
selection. A small “m” on the left side of the screen indicates the analyzer
has automatically jumped to a manual 25°C setting due to a failure with
the temperature signal input. Relay C indicator is also displayed here
(normally configured as FAIL system relay.)
32.0 nA
Raw sensor current. Useful for diagnosing problems.
100% 20.00 mA
PID Status screen (if enabled.) Shows the present controller output level
on left, and actual analyzer current on the right. The controller can be
placed in manual while viewing this screen by pressing and holding the
ENTER key for 5 seconds until a small flashing “m” appears on the
screen. At that point the controller output can be adjusted up or down
using the UP and LEFT arrow keys. To return to automatic operation,
press and hold the ENTER key for 5 seconds a
nd the “M” will disappear.
#14.00 mA
Analyzer output current # 1.
#2 12.00 mA
Analyzer output current # 2.
#3 20.00 mA
Analyzer output current # 3 (if option included.)
Aux relay=D,E,F
Auxiliary relay annunciators (if option included.)
Slope = 100%
Sensor output response vs. ideal calibration. This value updates after
each calibration. As the sensor ages, the slope reading will decay
indicating sensor aging. Useful for resolving sensor problems.
Offset = 0.0 nA
Sensor output current at a zero ppm input. This value updates after a
zero-calibration has been performed. This is useful for resolving sensor
problems.
Q46H0 v 4.02
Analyzer software version number.
7.56 pH
Measured pH value on AUX input (if enabled.)
814 mV
pH sensor mV output (if enabled)
Slope = 100%
pH sensor slope response vs. ideal calibration. This value will update
after each calibration. As the pH sensor ages, the slope reading will
decay. This is useful for solving sensor problems (if enabled).