7.3 CALIBRATION OF A/D GAUGE
x
Select and press SELECTOR to entry the A/D
gauge calibration program.
x
Turn SELECTOR TO select A or D shown on the
screen. The item highlighted in red is selected (ref.
fig. 54).
x
Press SELECTOR to make the calibration for the
item selected.
7.3.1 CALIBRATION OF A GAUGE
x
Select A by turning and pressing SELECTOR.
x
Place the gauge to the
“
0
”
position as shown in the
figure 55 .
x
Press SELECTOR to memorize the position.
x
Set the gauge at the distance 200mm as shown in
the figure 56.
x
Keep the gauge at 200mm distance and press
SELECTOR to memorize the position.
.
7.3.2 CALIBRATION OF D GAUGE
x
Select D by turning and pressing SELECTOR.
x
Place the gauge tip in lower hole of the standard
gauge of P4-40001 from the HIGER PROFILE
surface of shaft (
“
0
”
position) as shown in the figure
58.
x
Press SELECTOR to memorize the position.
x
Place the gauge tip in higher hole of the standard
gauge of P4-40001 from the HIGER PROFILE
surface of shaft (
“
200mm
”
position) as shown in the
figure 59.
x
Press SELECTOR to memorize the position.
x
Press STOP/ESC key to exit.
Fig. 54
Fig. 55
Fig. 56
Fig. 57
Fig. 58
Fig. 59