background image

34

12.4  Recommended separations between portable and mobile RF Communications 

         equipment and the ATMOS

®

 Aqua clean

Recommended separations between portable and mobile RF Communications

equipment and the ATMOS

®

 Aqua clean

The ATMOS

®

 Aqua clean is intended for use in electromagnetic environment in which radiated disturbances are 

controlled. The customer or user of the ATMOS

®

 Aqua clean can help prevent electromagnetic interference by 

maintaining a minimum distance between portable and mobile RF Communications equipment and the ATMOS

®

 

Aqua clean as recommended below, according to the maximum output power of the communications equipment.

Separation distance, depending on transmit-frequency m

Nominal output of the 

transmitter

W

150 kHz up to 80 MHz

d = 3,5/3√ P 

80 MHz up to 800 MHz

d = 3,5/3 √ P

800 MHz up to 2,5 GHz

d = 7/3 √ P

0,01

0,12

0,12

0,24

0,1

0,37

0,37

0,74

1

1,2

1,2

2,4

10

3,69

3,69

7,38

100

11,66

11,66

23,32

For transmitters for which the maximum nominal output is not indicated in the above table, the recommended 

separation distance 

d

 in meters (m) can be determined using the equation belonging to the respective column 

whereas 

P

 is the maximum nominal output of the transmitter in watts (W) acc. to manufacturer´s specification.

NOTE 1   With 80 MHz and 800 MHz the higher frequency range applies.

NOTE 2   These guidelines might not be applicable in any case. The propagation of electromagnetic sizes is  

                influenced by absorptions and reflections of buildings, objects and people.

NOTE 1     With 80 MHz and 800 MHz the higher frequency range applies.

NOTE 2     These guidelines might not be applicable in any case. The propagation of electromagnetic sizes is  

                  influenced by absorptions and reflections of buildings, objects and people.
a      The field strength of stationary transmitters, such as base stations of cellular phones and mobile terrain radio

 

        equipment, amateur radio transmitters, cbm broadcast and TV stations cannot be predestined exactly. 

        To determine the electromagnetic environment in regard to stationary transmitters, a study of the location is to 

        be considered. If the measured field strength at the location where the ATMOS

®

 Aqua clean is used exceeds 

        the above compliance level, the ATMOS

®

 Aqua clean is to be observed to verify the intended use.  

        If abnormal performance characteristics are noted, additional measures might be necessary, e. g. a changed  

        arrangement or another location for the device.

b      Within the frequency range of 150 kHz to 80 MHz the field strength is to be below 3 V/m.

12.0   Notes on EMC

Summary of Contents for 530.2121.0

Page 1: ...Gebrauchsanweisung ATMOS Aqua clean Water Supply System Operating instructions English GA1GB 110002 0 2018 05 Index 08 ...

Page 2: ...rigation and stimulation devices 11 4 0 Cleaning and disinfection 12 4 1 Disinfectant Changing the chemical bottle 12 4 1 1 Disinfectant 12 4 2 Care and cleaning 13 5 0 Maintenance 14 5 1 Period tests 14 5 2 Sending in the device 14 6 0 Retrofitting 15 26 6 1 Retrofitting at ATMOS S 61 Servant 15 20 6 1 1 Tubing 17 6 1 2 Retrofitting of a delay PCB 18 6 1 3 Delayed switching circuit for Aquastop 1...

Page 3: ...operating instructions must always be kept available near the device Care and period tests in conjunction with professional execution provide for operational safety and readiness for use of your ATMOS Aqua clean and are therefore a must besides regular cleaning Repair work and period tests may be carried out only by expert personnel authorised by ATMOS By applying only original spare parts you wil...

Page 4: ... ensures that water contaminated with micro organisms or mixed with chemicals cannot flow back into the water supply system If a water fed unit is activated the decontaminated water flows from the mixing container via the pressure container and to the appropriate instrument Naturally this causes the water level in the pressure container to fall together with the water pressure A pressure sensitive...

Page 5: ...rt 1 l ATMOS Green Clean WK connecting kit 1 5 Transport and storage Only transport the device in a shipping carton which is padded and offers sufficient protection If damage occurs during transport Document and report the transport damage Send the device to ATMOS see chapter 5 2 Sending in the device on page 14 Ambient conditions Transport Storage 5 40 C 05 95 air humidity non condensing at air p...

Page 6: ... 93 42 EEC REF SN Please observe operating instructions Proper disposal Manufacturer 1 6 Explanation of pictures and symbols 1 0 Introduction Short cuts symbols contained in this manual Please press where the dot indicates Subnumeration Numeration General information Follow the arrows whilst proceeding sequence Replace Check Please read important information Move plug in this direction Engage chec...

Page 7: ...t be liable for damage to people or property if Any non original ATMOS parts are used the user instructions given in this manual are not followed exactly or are disregarded Assembly resetting alterations extensions and repairs are not carried out by people authorised by ATMOS there is any damage in transport due to insufficient packing or due to residual water in the device Prior to starting up th...

Page 8: ...hapter 4 2 Starting Sequence 1 Open main water tap 2 Switch on ATMOS S 61 Servant 3 Switch on ATMOS Aqua clean If error F1 appears switch the ATMOS S 61 Servant off and on until the error disappears 3 2 Hose connections Chemical bottle connection PVC tubing Ø internal 2 mm Ø outer 4 mm Connection to the water outlet supplying the instruments PVC tubing Ø internal 4 mm Ø outer 6 mm Water connection...

Page 9: ...lly If this error was only produced by an air bubble it will expire within the next trial by itself However no decontamination takes place until the empty chemical bottle is replaced with a full one see chapter 4 2 1 The green control signal is illuminated and the red signal jitters and an alarm tone sounds Fig 8 Cause Level monitor in the mixing container is soiled or there is no water supply Act...

Page 10: ... clamp 061 0061 0 2 Hose barb 8 8 Legris 000 0974 0 3 Hose barb 6 5 Legris 000 0988 0 2 Fabric tube 6x3 1 m 005 0091 0 1 Fabric tube 8x3 1 m 005 0020 0 1 Bottle support 1 Seal 1 ATMOS Green Clean WK 534 3110 0 1 Starting Sequence 1 Open main water tap 2 Switch on ATMOS S 61 Servant 3 Switch on ATMOS Aqua clean If error F1 appears switch the ATMOS S 61 Servant off and on until the error disappears ...

Page 11: ...in water tap 2 Switch on irrigation and stimulation device 3 Switch on ATMOS Aqua clean If error F1 appears switch the irrigation and stimu lation device off and on until the error disappears 3 0 Setting up and starting Chemical bottle connection PVC tubing Ø internal 2 mm Ø outer 4 mm Connection to the water outlet supplying the instruments PVC tubing Ø internal 4 mm Ø outer 6 mm Water connection...

Page 12: ...tion can be withdrawn The empty bottle should be detached from the fitting tube on the ATMOS Aqua clean and replaced with a new one Please note that only ATMOS Green Clean WK provides the concentration required for broadspectrum disinfection The use of chemicals other than from ATMOS may damage the ATMOS Aqua clean as well as other devices which are connected Safety precautions Keep out of the rea...

Page 13: ... signal is activated The ATMOS Aqua clean returns into normal operation Intensive decontamination is always recommended prior to longer idle times like weekend or holiday and after treatment of patients with suspicion of HIV HBV or HCV Remove water level probe from the mixing container and clean or replace Switch off main switch Carefully pull probe up and out Clean or replace probe Check the elec...

Page 14: ...eriod tests Functional and tidiness inspections are required after any work at the ATMOS Aqua clean At least every 12 months a repeat test of the electrical safety should be performed according to IEC 62353 ATMOS recommends an inspection according to the manufacturer s specifications 5 2 Sending in the device Remove and properly dispose of consumables Clean and disinfect the product and accessorie...

Page 15: ...8 1 8 Legris 000 0976 0 1 piece Hose barb 8 8 Legris 000 0974 0 3 pieces Screw in union straight Rd6 1 8 Legris 000 0975 0 1 piece Hose barb 6 5 Legris 000 0988 0 2 pieces Screw in union L Rd8 1 8 Legris 000 0981 0 1 piece Mounting plate 530 2122 0 1 piece Valve mounting angle 530 2123 0 1 piece Disk 22 x 16 x 1 053 0113 0 2 pieces Disk 22 x 18 x 1 053 0114 0 2 pieces Hose quick connector T 4 4 4 ...

Page 16: ...Exchange existing plate against mounting plate 15 16 Please drill the required holes into ther rear part of the housing as per the drawing Starting Sequence 1 Open main water tap 2 Switch on ATMOS S 61 Servant 3 Switch on ATMOS Aqua clean If error F1 appears switch the ATMOS S 61 Servant off and on until the error disappears diam 18 mm drill with hole cutter diam 16 mm drill with hole cutter 4 5 6...

Page 17: ...005 0020 0 1000 mm Hose Rilsan RD A 8 0 005 0025 0 240 mm Hose Rilsan RD A 6 0 005 0173 0 200 mm Hose Rilsan RD A 4 0 005 0026 0 450 mm Hose Rilsan RD A 4 0 005 0026 0 250 mm Hose Rilsan RD A 4 0 005 0026 0 680 mm Hose Rilsan RD A 8 0 005 0025 0 130 mm 1 2 3 4 5 6 7 1 2 3 4 5 6 7 Fig 24 8 8 water connection non return valve ...

Page 18: ...e and mark the drillings Afterwards drill the holes with a 3 5 mm drill Screw the 4 bolts M3x10 including washer and locknut Tighten the PCB with 4 lens head screws 3x5 Fix the connecting cable REF 530 2124 0 as per the drawing Connect the existing Aquastop to the PCB REF 011 1314 0 6 0 Retrofitting Fig 25 Fig 26 ...

Page 19: ...B for 230V switching 230V Control PCB ST1 011 1307 0 Aquastop Verzögerung Delay for Aquastop switching Verzögerungsschaltung für Aquastop S 61 Servant Delayed switching circuit for Aquastop on ATMOS S61 servant spülbeh sekretbeh reserve Steuerplatine Control PCB 530 1149 0 S5c S5 530 2124 0 12V GND ...

Page 20: ...000 0978 0 Ursprünglich am T Stück Originally on T Connector Druckerhöhung auf 3 5 bar Pressure elevation to 3 5 bar Schlauch RILSAN 8mm Hose RILSAN 8mm 005 0025 0 Schlauch RILSAN 8mm Hose RILSAN 8mm 005 0025 0 Schlauchverbinder L 8 8 Hose Connector L 8 8 000 0252 0 Ohrspülmodul Ear Irrigation Module Manometer Schlauch RILSAN 6mm Hose RISAN 6mm 005 0173 0 Schlauchverbinder 1 8 6 Hose Connector 1 8...

Page 21: ...0 0252 0 1 piece Hose Rd I 6 0 x 3 0 005 0091 0 1000 mm Hose Rd I 8 0 x 3 0 005 0020 0 1000 mm Hose Rilsan Rd A 4 0 005 0026 0 1000 mm Hose Rilsan Rd A 6 0 005 0173 0 200 mm Hose Rilsan Rd A 8 0 005 0025 0 1000 mm Bottle support 1 piece Seal 1 piece ATMOS Green Clean WK 534 3110 0 1 bottle 6 2 1 Drilling Note When drilling the holes for the screwed connection water in outlet into and out of the AT...

Page 22: ...22 6 0 Retrofitting 11 10 3 1 7 12 5 2 6 9 4 8 1 Fig 29 Fig 30 6 2 2 Tubing 6 2 3 Add on kit heating module existing new ...

Page 23: ...ST1 ST2 KL1 Steuerplatine Servant 5 Control PCB Servant 5 510 0040 0 011 1307 0 Aquastop Verzögerung Delay for Aquastop switching PCB Verzögerungsschaltung für Aquastop Servant 5 bis Bj 2005 Delayed switching circuit for Aquastop on Servant 5 until 2005 Ursprünglicher Anschluß Aquastop Original Aqustop connection ST15 ST11 Vorhandenes Kabel Stirnlampenhaken Existing cable for headlamp suspension 5...

Page 24: ...eckenspülung Bowl rinsing Pegelschalter Level Switch Schalter Schlauchspülbehälter Microswitch Hose Rinsing Bottle Ohrspülmodul Ear Irrigation Module Fallrohrelektroden Drain Pipe Electrodes Füllstandselektroden Sekretbehälter Level Electrods Secretion Jar Taster Spiegelschnellerwärmer Push Button Mirror Quick Heater Sicherheitsschalter Abdeckklappe Cover switch Steuerplatine Control PCB Servant 5...

Page 25: ...asserfilter Water Filter 502 0890 0 Schlauch RILSAN 8mm Hose RILSAN 8mm 005 0025 0 Schlauch RILSAN 8mm Hose RILSAN 8mm 005 0025 0 Druckminderer Pressure reducer Schlauch RILSAN 6mm Hose RILSAN 6mm 005 0173 0 Schlauchschnellverbinder I 1 8 6 Hose Quick Connector I 1 8 6 Anschluß Zulauf Heizblock Ohrspülung Inlet connectot heating block ear irrigation Wasserdruck Water pressure Wasserzulauf Water In...

Page 26: ...htülle 8mm Spout 8mm 000 0974 0 Schlauchtülle 6mm Spout 6 mm 000 0988 0 Schottverschraubung 6 6 Bulkhead connector 6 6 000 0980 0 Schottverschraubung 8 8 Bulkhead connector 8 8 000 0978 0 Druckminderer Pressure Reducer 502 0955 0 Schlauch RILSAN 6mm Hose RISAN 6mm 005 0173 0 Schlauchverbinder 1 8 6 Hose Connector 1 8 6 000 0975 0 Schlauch RILSAN 4mm Hose RILSAN 4mm 005 0026 0 Schlauchverbinder 1 8...

Page 27: ... you follow these notes Error Possible reason Troubleshooting Device pump does not start up Loose power plug Check connection to supply socket No mains voltage Check inbuilding fuse Check whether power plug on the unit is loose Defective fuse at the power input Exchange fuse 8 0 Accessories and Spare Parts Article description REF ATMOS Green Clean WK 534 3110 0 Service kit 069 0260 0 ...

Page 28: ...ner here water is mixed with the disinfectant after the cascade section This module includes the container filling and overflow sensor 2 magnetic valves and the disinfectant sensor Module 5 Double membrane pump and membrane break sensor Module 6 Water unit consists of a flow rate limiter and dosing valves Module 7 Compressed air unit this is a pressure regulator Module 8 Auxiliary installation hou...

Page 29: ... valve on the pressure container black MV4 External display Chemical valve black MV3 Water safety valve 5 brown Pump motor connection Membrane breaking yellow Power supply plug 24 V AC orange Water dosing valve 1 brown Compressed air valve black MV2 Main fuse 3 15 A T When replacing the fuses always use the same type 9 3 Wiring diagram see Fig 39 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 1 2 3 4 5 6 7 8...

Page 30: ...d water pressure 2 6 bar Operating pressure water 2 2 bar Operating pressure air 4 bar Maximum water flow rate 550 ml min Water quality drinking water Disinfectant ATMOS Green Clean WK Hydrogen peroxide 2 solution Mixing ratio 1 85 standard setting 1 42 intensive decontamination Working solution 235 ppm Overall dimensions H x W x D 550 x 265 x 160 mm Fig 40 ...

Page 31: ...cycle it must be disposed of professional According to the EC directives WEEE and RoHS the device may not be disposed of in domestic waste Please observe existing national laws and rules for disposal of old devices Disposal within the Federal Republic of Germany In the Federal Republic of Germany the law for electrical devices ElektroG rules the disposal of electrical devices In order to guarantee...

Page 32: ...gs used for domestic purposes Harmonics IEC 61000 3 2 Class A Flicker IEC 61000 3 3 Compliance 12 2 Guidelines and Manufacturer s Declaration Immunity for ATMOS Aqua clean The ATMOS Aqua clean is intended for use in the electromagnetic environment specified below The customer or user of the ATMOS Aqua clean should ensure that it is used in such an environment Immunity Test IEC 60601 Test Level Com...

Page 33: ...Guidelines and Manufacturer s Declaration Immunity The ATMOS Aqua clean is intended for use in the electromagnetic environment specified below The customer or user of the ATMOS Aqua clean should ensure that it is used in such an environment Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance Conducted RF IEC 61000 4 6 3 Veff 150 kHz to 80 MHz 3 V Portable and m...

Page 34: ...nominal output of the transmitter in watts W acc to manufacturer s specification NOTE 1 With 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines might not be applicable in any case The propagation of electromagnetic sizes is influenced by absorptions and reflections of buildings objects and people NOTE 1 With 80 MHz and 800 MHz the higher frequency range applies NOTE 2 Th...

Page 35: ...35 For your notes ...

Page 36: ...ATMOS MedizinTechnik GmbH Co KG Ludwig Kegel Straße 16 79853 Lenzkirch Germany Phone 49 7653 689 0 atmos atmosmed de www atmosmed com ...

Reviews: