32 Cleaning and disinfection
5 Cleaning and disinfection
We recommend that you always document all maintenance work and exchange of parts
in writing.
5.1 Cleaning and disinfection
Risk of infection.
Cross infection is possible. Diseases could be transmitted.
•
Clean and disinfect the device according to the operating instructions.
•
Always wear disposable gloves during any cleaning.
•
Clean the device and accessories after each use.
•
Observe the valid guidelines.
Material damage due to liquids or wrong disinfection.
•
The cleaning cloth must not be wet.
•
Never spray disinfectant directly onto the device or accessories, always on the cloth.
•
Do not rinse the device or accessories under running water and do not immerse it
into any liquids.
•
Do not autoclave the device.
•
Take care that no liquid penetrates the device. Pay particular attention to the con-
nections. In case that liquid has penetrated the device it may no longer be operated.
In this case clean the device and send it in to ATMOS for repair.
•
Observe the operating instructions of the disinfectant manufacturers. Note in
particular the information on the concentration and material compatibility.
•
Only use disinfectants, which are recommended by ATMOS (chapter „5.3 Recom-
mended disinfectants“ on page 33). The use of other disinfectants may damage
the device.
ATMOS
®
Audio Cube
Clean the ATMOS
®
Audio Cube after each patient.
Electric shock due to liquid in the device.
1. Disconnect the device from the power supply.
Porous plastic due to detergents containing solvents.
2. Dampen a soft cloth with water and a mild, solvent-free detergent.
3. Clean the device with a damp (not wet) cloth.
4. Disinfect the housing with a recommended disinfectant.
5.
Make sure the ATMOS
®
Audio Cube is dry before you switch it back on.
Headphones, bone vibrator and patient response button
Replace the ear cushions and headband pads at regular intervals.
Clean the headphones, bone vibrator and patient response button after each patient.
CAUTION
ATTENTION