40 Maintenance and service
6.5.4 Bone vibrator cable B71
1. Switch off the device.
2. Loosen the cable of the bone vibrator from the audiometer.
3. Gently pull the cable plug from the bone vibrator capsule.
4. Plug in the new cable to the bone vibrator capsule.
5.
Plug in the new bone vibrator cable into the audiometer.
6.5.5 Mains fuses
Fire hazard due to unsuitable fuses.
Burns and injuries are possible.
•
Only use glass tube fuses type T1A L/250V 5x20 mm.
•
Always replace the fuses in pairs.
•
If the fuses blow within a short period of time then an authorized service partner
must be contacted immediately.
1. Switch off the device.
2. Remove the power cable from the device.
3. Loosen the top and bottom of the fuse holder with a screwdriver.
4. Remove the fuse holder.
5.
Replace both glass tube fuses.
6.
Reattach the fuse holder. Press until it audibly locks.
6.6 Sending in the device
1. Remove and properly dispose of consumables.
2. Clean and disinfect the product and accessories according to the operating instruc-
tions.
3. Place used accessories with the product.
4. Fill in the form QD 434 „Delivery complaint / return shipment“ and the respective
decontamination certifi cate
.
&
This form is enclosed with each delivery and can be found at www.atmosmed.com
.
5.
The device must be well padded and packed in suitable packaging.
6.
Place the form QD 434 „Delivery complaint / return shipment“ and the respective
decontamination certifi cate
in an envelope.
7. Affi x the envelope to the outside of the package.
8. Send the product to ATMOS or to your dealer.