8 Hints for your safety
2 Hints for your safety
The safety of the ATMOS
®
Audio Cube 31 complies with all the recognized rules of
technology and the directives of the Medical Devices Act.
Read and follow the safety information carefully before using the product.
General safety information
If you wish to operate the ATMOS
®
Audio Cube 31 in combination with non-medical
products, (e.g. PC, printer), observe the chapter „3.5 Use with other devices“ on page
12.
The device has application parts from type B.
Dangers for users, patients and third parties
Protect yourself against an electric shock!
Burns, cardiac arrhythmias and even death are possible.
•
Prior to each use check whether the device or power cable are damaged. Never
operate the unit, if it shows any obvious safety defects. In this case clean the device
and send it in to ATMOS for repair.
•
Disconnect the device from the mains power supply prior to cleaning or disinfection.
•
Prior to switching the device on/off remove the transducer from the patient or
examiner.
•
Plug out the transducer before you repair it. First remove the transducer cable from
the audiometer before you loosen the cable from the transducer itself.
•
Do not operate the device if it has been dropped. In this case clean the device and
send it in to ATMOS for repair.
•
Only connect the device to a power supply with a correctly installed protective
conductor.
•
Only use accessories and spare parts from ATMOS.
•
Assembly, new settings, alterations, extensions and repairs may only be carried out
by authorized persons.
•
Pay attention to the period tests in chapter „6.1 Period tests“ on page 35.
Explosion and fire hazard!
Burns and injuries are possible.
•
Never operate the product ATMOS
®
Audio Cube 31 in explosion-hazardous areas or
areas which are oxygenated.
•
Only use original accessories and original spare parts from ATMOS.
Reduce the risk of infection for you and your patients!
Diseases could be transmitted.
•
Clean and disinfect the product as described in chapter „5 Cleaning and disinfection“
on page 32.
•
Use standard protective covers for headphones.
•
Replace the ear cushions and headband pads at regular intervals.