19
4.3.3
Secretion canister evacuation,
fully-automatic
(option)
Purpose:
•
Prevents interruption of work in practice.
•
Secretion is not held in canister over a longer period of
time.
•
Prevents contact of staff with secretion.
Activated automatically when:
• Reaching the upper fi lling limit (electrodes).
•
Returning suction attachment to holder.
The tube cassette of the tube pump is a wear part and
must be replaced at regular intervals (see chapter 6.2) !
Pay attention to cleaning described in chapter 5.0 !
4.3.4
Quiver for cleaning and
disinfecting solution
(option)
Cannulae and suction hose should be rinsed through each
time after use to prevent sticking and clogging up. We
therefore offer an optional quiver that may be attached to
the handle support (see fi g. 20).
After cleaning, cannulae may no longer be used in patients
(exception: if the cleaning solution is changed after every
patient).
Particularly suitable: ATMOS special cleanser (080.0005.0).
Fig. 19. Service compartment
Connecting nipple
Secretion canister
Hose rinsing container (option)
Secretion hose
Hose for automatic secretion canister evacuation
Connecting electrodes for automatic
secretion canister evacuation (optional)
Connection for secretion hose
Fig. 20. Plastic quiver for cleaning and
disinfecting solution (506.7015.0)
4.3.2
Mirror preheater
(option)
The mirror preheater is switched on when the main switch
(
, fi g. 1, page 7) is activated.
The rack (
, fig. 18) containing the mirrors is
heated up to about 40°C.
ENT unit should be switched on about 20 minutes
before starting work, to ensure preheating in good
time!
Attention, high temperatures !
Prior to use of the mirror, always check its temperature
(with hand etc.) !
4.0 Operation