25
5.1
General information on cleaning and
disinfection
Set main switch to OFF prior to cleaning and
disinfection!
The described measures for cleaning do not re
place the respective precautions for operating the
unit !
For disinfection, you may use all surface and instrument
disinfectants listed on page 26.
Always observe the concentration specifi cations and
instructions by the respective manufacturer !
Do not use
• Disinfectants containing concentrated organic or an-
organic acids or bases, since these may cause
corrosion damages;
• Disinfectants containing chloramides, phenol deri-
vatives or anionic tensides, since these may cause
stress cracks in the material used for the housing of
the unit.
5.1.3
Secretion canister (without autom.
secretion canister evacuation),
bacterial fi lter and suction hose
At the end of every working-day,
following parts must be
cleaned and disinfected:
Secretion canister with lid system and bacterial fi lter:
•
Detach all hose connections carefully on the lid system
and take secretion canister out carefully to prevent spills and
contamination of the area. Dispose of secretions
properly.
• Grip lid system fi rmly, open lid of fi lter housing by turning
in anti-clockwise direction.
• For systems with bacterial fi lter plates or integrated
bacterial fi lter:
Remove fi lter and dispose of it.
• With external bacterial fi lter:
To open the bacterial fi lter cover please turn it. Remove
fi lter and dispose of it.
• After cleaning a new filter must be inserted (with
bacterial fi lter plate: smooth side down).
See chapter 6.1.
• The bacterial fi lter is a disposable and must be disposed
of. The bacterial fi lter must be exchanged at least once
a day.
Suction system and hose attachment:
• After every use, rinse out the suction system by
drawing in a small amount of irrigating fluid (e.g.
ATMOS Special Cleanser 080.0005.0).
Keeps the hoses from becoming sticky or
clogged.
Suction capacity is limited by the 1.25 l secretion
canister. Therefore, do not use more than 1 l rinsing
liquid and subsequently evacuate the canister.
Replace The fi lter.
5.1.2
Cleaning "application parts"
"Application parts" comprise:
All single components or assemblies which come into
contact with the patient and might get contaminated:
•
secretion canister,
•
secretion hose,
•
nozzles of medicament sprayers,
•
Politzer olives and adapters.
All application parts may disinfected with instrument
disinfectants (see page 26).
All application parts which are exposed to direct
contact with the patient during treatment are to be
exchanged or cleaned immediately for hygienic
reasons.
5.1.1 Cleaning the unit surface
The surfaces of the ATMOS
®
C 31
are resistant against all
listed surface disinfectants.
Wipe the unit surface with a cloth moistened with a cleaning
or disinfecting solution.
Wipe dry the device surface, the surface edges may not
be wet for a longer period.
5.1.4
Secretion canister and electrodes
(automatic evacuation of
secretion canister)
The secretion canister should be removed and cleaned once
every week.
•
During the cleaning of the canister the cover has to be
pulled off and rinsed thoroughly under running water.
Use the disinfectants mentioned in chapter 5.4 for
disinfection. Before the canister is assembled again, the
electrodes in the cover should be cleaned using a wet
cloth.
Mistake in polarity not possible!
5.0 Cleaning