11.0 Notes on EMC
)
Medical electrical equipment is subject to special precautions with regard to EMC
and must be installed according to the following EMC notes.
Guidance and manufacturer’s declaration – ambient conditions
The ATMOS HL 21 LED is suitable for use in the following environments:
•
In fields of home health care in any buildings, outdoor areas, and means of
transport.
•
In professional healthcare facilities such as medical practices, hospitals/clinics, first-
aid facilities, and operating theatres/rooms.
The following environments are not suitable:
• Vicinity of HF surgical devices and in settings outside of an HF-shielded room of a
magnetic resonance imaging system.
• Special environments such as factory or military facilities and medical areas near
HF surgical devices, short-wave therapy equipment, or within an HF-shielded room
of a magnetic resonance imaging system.
The customer or user of the ATMOS HL 21 LED must ensure that it is used in a
prescribed environment.
Guidance and manufacturer’s declaration – key features
)
Please note the Technical Data in these instructions. The essential features are fully
usable even in the presence of electromagnetic disturbances.
Guidance and manufacturer’s declaration – removable components that can be
replaced by the operator
The ATMOS HL 21 LED
has the following removable components that can be replaced
by the operator:
Type
REF
Max. cable length
Charging power pack
011.1199.0
1.8 m
Connecting cable
507.4542.0
0.4 m
Connecting cable
507.4541.0
1.4 m
Connecting cable
507.4545.0
2.0 m
Guidance and manufacturer’s declaration – warnings
WARNING
The use of electrical components and accessories other than those specified or
provided by the manufacturer may cause increased electromagnetic interference or
reduced immunity to electromagnetic interference and result in faulty operation of the
device.