Target Point Calibration 

Note: Calibration needs to be done by professionals and with necessary accessories and standard gas. 

 

Prepare the standard gas before starting the target point calibration. In detection mode, press 

“MENU”, select 

“calibrate span”, press ON/OFF button to enter. Then you will need to enter password. (If you need password, 
please contact the manufacturer to get it) 

After entering the password successfully, press ON/OFF to enter the target point calibration interface, as 

shown in the picture. The interface will display 

“concentration”, “set value. The concentration is the result got 

from the calculation of several internal calibration parameters. The process of calibration is the process to 

recalculate the internal parameters of the device. The set value is the value need to be entered manually; it 

should be the concentration of the standard gas to be calibrated. Select the parameter under the current 

interface, modify its value with 

“PLUS” and “MINUS” button.   

 

The calibration process is as below: 

1.  Connect the detector with standard gas cylinder by hose. 

2.  Enter the target point calibration interface, set the 

“SET value” as the standard gas concentration. 

3.  Release the standard gas to the detector in a flow rate of 500mL/min. The value of 

“concentration” will 

become larger gradually, and become stable after 30seconds. 

4.  When the value of 

“concentration” is stable, press the “ON/OFF” button, the value of “concentration” will be 

same with or almost same with the 

“Set value”. The calibration is finished. 

5.  Cut off the standard gas supply. Press 

“ESC” to go back to detection interface. 

 

 

Summary of Contents for SKY2000

Page 1: ...e pump Power On Off Put the detector in clean air and press the ON OFF for three seconds the detector will be turned on and the pump will start to work The screen displays the following interface successively Brand and Logo main parameter interface initialing countdown The countdown would be 60 or 120 seconds which is to make sure that the sensors are fully activated After countdown the device ent...

Page 2: ...then observe the reading if it keeps going down then wait until the reading back to zero before powering it off and do the zero point calibration next time before using it If the reading maintained at full scale user should send the device back to manufacturer or agent for repair or replacing the sensor Air Inlet Air Outlet air chamber of sensor LCD PLUS MINUS MENU ESC ON OFF Note If detector is u...

Page 3: ...light will be flickered Alarm Setting Press Menu choose the Alarm point set and start to set the alarm value There are LOW POINT and HIGH POINT for set press MENU to move the cursor to the value you want to set and press the PLUS or MINUS to change the value After the setting is done press ON OFF button to save the settings Low Battery Alarm When the battery is low there will be sound light low ba...

Page 4: ...n parameters The process of calibration is the process to recalculate the internal parameters of the device The set value is the value need to be entered manually it should be the concentration of the standard gas to be calibrated Select the parameter under the current interface modify its value with PLUS and MINUS button The calibration process is as below 1 Connect the detector with standard gas...

Reviews: