background image

7

January 2004

GPT ARGUS

Cleaning the ARGUS Validator

Depending on its environment and amout of use, the

Argus Bill Validator may require periodic cleaning to

restore the unit to optimum performance. Under nor-

mal use, it should be cleaned every 6 month. For

unusual operating conditions more frequent cleaning

will be required.

1. Open Main Door and turn power off.

2. Slide the Bill Validator Tray out.

3. Disconnect the connector cable and remove the

Bill Validator Head from its mounting frame.

4. Open the Bill Validator Head by releasing the

spring-loaded latches and swinging out the Up-

per-Guide Assembly.

5. Use a soft cotton cloth dampened with 90% iso-

propyl alcohol to carefully clean all parts in the

Bill Channel. Remove all deposits from the rollers.

Note

: Do not scrach any area, as this can dam-

age the Validator.

- Do not allow liquid to enter the Vallidator Unit, as

this can damage electrical components.

- Do not use unapproved cleaners, as this can dam-

age the surface of the optics.

- Do not use cotton swabs, as this can leave un-

wanted material in the Bill Channel.

6. Carefully swing the Upper-Guide Assembly back

to its fully closed (locked) position. Make sure no

cable is squeezed by the Upper Guide Assembly.

7. Mount the Bill Acceptor Head back into its mount-

ing frame and connect the connector cable

8. Slide the Bill Validator Tray in. Make sure that it

is clamped in its position.

9. Apply power and close the Main Door. The Argus

Bill Acceptor is now operational.

The use of  Validator Cleaning Cards is not recom-

mended. These cards need to rub against the optic

and magnetic head in order to clean the components.

In the ARGUS validator the optics are recessed and

therefore cannot be reached by the Cleaning Card.

Note

: Do not squeeze

cables when closing the

Validator Head.

V

ALIDATOR

 H

ANDLING

Summary of Contents for GPT ARGUS B

Page 1: ...www atronic com e motion Hi bility Bill Validator GPT ARGUS B D Rev 1 2 January 2004 ...

Page 2: ...2 January 2004 GPT ARGUS Bill Validator GPT ARGUS B D Rev 1 2 Rel January 2004 2003 2005 Atronic All rights reserved RELEASE INFO ...

Page 3: ...ance via host machine Easy opening spring latches design Upper Guide Assembly Front Optical Sensor Pressure Rollers 6 total Mag Head Roller Side Looking Sensor Main Optics Window lower Drive Rollers 1 2 3 4 5 6 7 ARGUS B Validator Head Back Stack ARGUS D Validator Head Down Stack Enclosure for ARGUS Head ARGUS D G3 Validator Head opened 22 Pin Connector DIP Switches Drive Belts Rear Optical Sensor...

Page 4: ... motor and draws the bill into the channel b The bill passes the main optics window and the mag head c The CPU collects and analyses the optical and magnetic data d The collected data is then compared with a known data base to determine validity and denomination of the bill 3 Acceptance Routine a If the bill is determined to be valid the CPU con tinues in the sensor checking mode b The rear channe...

Page 5: ... bill present These decisions may be influenced by the programmed security level The level will generally have no effect on the acceptance of bills in good condition but may reduce the acceptance by highly circulated bills Technical Information Inspection 1 Ensure the unit has not been dented or bent check for squareness and all mounting surfaces are free from damage 2 Check that there are no brok...

Page 6: ...pply a RS 232 cable a ARGUS Download Key and the Synergy G3comm2000c Software You can also test IO ports memory stepper motor and bezel lights with this software Please contact Atronic Technical Service for further information Maintenance Most of the problems with the validators are caused by soiled optics incorrect DIP switch settings and foreign matter in the currency channel Component Checks Dr...

Page 7: ...am age the Validator Do not allow liquid to enter the Vallidator Unit as this can damage electrical components Do not use unapproved cleaners as this can dam age the surface of the optics Do not use cotton swabs as this can leave un wanted material in the Bill Channel 6 Carefully swing the Upper Guide Assembly back to its fully closed locked position Make sure no cable is squeezed by the Upper Gui...

Page 8: ...led VLA cards for this test Please ask Atronic Technical Service to get a new matching VLA card VLA procedure Note If a software update needs a hard boot per form this procedure directly after software upload 1 Power down the Validator 2 Set DIP switch 8 9 to the ON down position 3 Power up the Validator 4 Wait approx 5 seconds for Side Looking Sensors SLS to readjust The Status LED will illuminat...

Page 9: ...e or the rejecting state All green LEDs are not lit The red LED is lit Bezel LED displays error codes Error Malfunction 1 A bill is inserted in the channel but does not advance into the channel 2 Stacker is not connected 3 Validator is in Inhibited Mode caused by an inhibit command from the controller Green LEDs are not lit Red LED is lit Bill jams in the channel Middle Sensor is covered Front Sen...

Page 10: ...i e r o F l e n n a h c e h t e v i r d d e s o p x e e h t n a e l C e r u s s e r p d n a s t l e b e v i r d k c e h c s r e l l o r d n a e g a m a d r o f s t l e b n o i s n e t r e p o r p r o f s t c e j b o n g i e r o f e v o m e R e r u s n e l e n n a h c e h t m o r f l l a f o e e r f s i l e n n a h c s i r b e d e k c a t S r s n o i t c n u f l a m s i t o l s y r t n e r e k c a ...

Reviews:

Related manuals for GPT ARGUS B