To Find out how many Months are left on the Lamp Timer -
Push the reset button for 1-2 seconds
then release. After an initial sound there will be a pause followed by beeping and a flashing LED. The
number of beeps corresponds to the number of months left on the lamp life.
To Reset 1 Year Lamp Timer
– WHEN REPLACING THE UV LAMP. Unplug the control box from
the wall outlet. Unplug the UV lamp from the Lamp connector. Replace with a new UV Lamp. Make
sure the control box is plugged into the wall outlet and the UV lamp is on. Press and hold the reset push
button. A tone that descends in frequency once per second will be heard. Hold the reset button down for
approximately 5-8 seconds until you hear a continuous tone. At this point release the button. The 1 year
timer has just been reset.
LAMP STATUS INDICATOR / UV LAMP ADVISOR:
LED
COLOR
AUDIBLE
ALARM
UNIT
STATUS
GREEN
NONE
OK
FLASHING
YELLOW
INTERMITTENT
TONE
REPLACE LAMP WITHIN
30 DAYS.
FLASHING
RED
INTERMITTENT
TONE
LAMP HAS EXCEEDED 1 YEAR
OF OPERATION, REPLACE
LAMPS & RESET.
CONTINUOUS
RED
CONTINUOUS
TONE
UV LAMP IS NOT LIT.
NOTE: During normal operation, the Green LED will flash every 5 - 10 seconds.
Please contact your local dealer for replacement UV lamps.
Refer servicing to qualified service personnel.
Summary of Contents for DWS-15
Page 6: ...6 ATS 280CB RESET BUTTON MOUNTED UNDER TOP PLATE OUTLET MAINS BALLAST LAMP...
Page 19: ...19...
Page 20: ...20 WIRING DIAGRAM...
Page 21: ...21 ATS1 421 CCASE WH 3C DWS 15 DWSW 15...
Page 22: ...22 ATS1 421 CCASE WH 3C...
Page 23: ...23...
Page 24: ...24 ATS1 421 WH 3C...
Page 25: ...25...
Page 26: ...26...
Page 27: ...27 DRAWING AND PART NUMBERS...
Page 28: ...28 DWS 7 Explosion...
Page 30: ...30 DWS 15 Explosion...
Page 32: ...32 SE 7 Explosion...
Page 34: ...34 SE 15 Explosion...
Page 36: ...36 DWSW 8 Explosion...
Page 38: ...38 DWSW 15 Explosion...