an immediate offset measurement.
[SENSe:]GAIN:AUTO?
Query the auto gain mode. Returns “1”(ON) or “0”(OFF or ONCE).
INPut:IMPedance:AUTO{OFF|ON}
Disable or enable the automatic input resistance mode for DC voltage
measurements. With AUTO ON, the input resistance is set to >10GΩ for the
100mV, 1V and 10V ranges. With AUTO OFF, the input resistance is fixed at
10MΩ for all ranges.
INPut:IMPedance:AUTO?
Query the input resistance mode. Returns “1”(ON) or “0”(OFF).
ROUTe:TERMinals?
Query the multimeter to determine if the front or rear input terminals are
selected. Returns "FRON" or "REAR"
ROUTe:CLOSe <channel>
Set channels which need to be closed. <The range is from channel 1 to 10>
ROUTe:CLOSe?
Query channels which were closed.
ROUTe:OPEN
Open all channels.
ROUTe:STATe?
Read the state after scanning. The state display is
1
(means the scanner card
was inserted.), or
0
(means it hasn’t been inserted yet.)
ROUTe:SCAN:FUNCtion
<channel>,{<function>|“VOLT:DC”|“VOLT:AC”|
“FREQuency”|“RESistance”|“FRESistance”|“NONE”}
Set scanner card’s measurement functions which might be for the VAC, VDC,
Frequency, 2-Wire Resistance, 4-Wire Resistance or disabling the channel.
163