an immediate internal trigger, or a hardware trigger from the rear-panel EXT
TRIG terminal.
TRIGger:SOURce?
Query the trigger source.
TRIGger:DELay {<
seconds
>|MINimum|MAXimum}
Set a trigger delay time in seconds. The delay is the time between the trigger
signal and each sample that follows. Specify a delay time from 0 to 3600
seconds.
TRIGger:DELay?
Query the trigger delay time.
TRIGger:DELay:AUTO {OFF|ON}
Disable or enable a automatic trigger delay. The delay is determined by
function, range, integration time, and ac filter setting. Specifying a delay time
automatically turns off the automatic trigger delay.
TRIGger:DELay:AUTO?
Query the automatic trigger delay mode. Returns “0” (OFF) or “1” (ON).
SAMPle:COUNt {<
value
>|MINimum|MAXimum}
Set the number of readings (samples) the multimeter takes per trigger. Select
a number from 1 to 50,000 readings per trigger.
SAMPle:COUNt ? [MINimum|MAXimum]
Query the sample count.
TRIGger:COUNt {<
value
>|MINimum|MAXimum|INFinite}
Specify the number of triggers the multimeter will accept before returning to
the “idle” state. Select from 1 to 50,000 triggers. The INFinite parameter
instructs the multimeter to take triggers continuously. Trigger count is ignored
while in local operation.
TRIGger:COUNt? [MINimum|MAXimum|INFinite]
Query the trigger count. Returns “9.9037” when you select the