ENTER button.
4
Press CONT button.
5
The measured value will be shown on the display automatically.
Meter “beeps” when measured resistance value is lower than
threshold value.
Figure 3-10
3.6 Diode Measurements
ATM3500A uses a current source of 1 mA for diode testing. The
maximum resolution is 10 μV on a fixed range of 1 V DC. The default
threshold voltage is fixed between 0.3 and 0.8 volts and the reading
rate is fixed at 0.1 PLC (The voltage bound is adjustable from 0.01V up
to 1.2V.). A “Beep” sound will appear when the diode measured value
is in the range.
Warning! Positive end of the source signal is to be connected to HI
of the input terminals, and the negative end to LO of the input
terminals.
How to measure diode
1
Selects input signal connections on front or rear panel.