constants required for that subrange are listed.
Default
The default sensor type, units and transducer in ATM3500A is PT100,
and 4W RTD in order.
℃
How to set up RTD
You can set up the RTD configuration either through the front panel
operation or through the remote interface operation.
Front Panel Operation
If you are using RTD, press CONFIG and then TEMP. Use
◁
and
▷
to
locate SENSOR submenu. Press ENTER to go to the submenu. Use
◁
and
▷
to locate your sensor type Press ENTER to select your desired
sensor type. So are the Units and Transducer settings.
Choosing
USER
takes you to a menu where you can specify factors
used in the calculation equation to obtain the temperature. Use
◁
and
▷
to move through the digits and
△
and
▽
to change the numbers to a
desired value. Press ENTER to set the value.
Choosing
SPRTD
takes you to a menu where you can specify the
seven coefficients
that are used to determine the temperature. Use
◁
and
▷
to move through the digits and
△
and to change the numbers
▽
to a desired value. Press ENTER to set the value.
Remote Interface Operation
Use the following commands to set up the RTD configuration.
SENSe:TEMPerature:RTD:TYPE{PT100|D100|F100|PT385|PT3916|USER|SPRTD|NTCT}
SENSe:UNIT {Cel|Far|K}
SENSe:UNIT?
SENSe:TEMPerature:RTD:RZERo {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:RTD:ALPHa {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:RTD:BETA {<
value
>|MINimum|MAXimum}
SENSe:TEMPerature:RTD:DELTa {<
value
>|MINimum|MAXimum}