Try running an On-ear test.
No test box reference mic. detected
No test box reference mic. detected!
This message may be displayed when starting a hearing instrument test. It indicates that
the Verifit is not measuring any sound at the test box reference microphone. The test box
reference microphone is the white microphone located in the test chamber.
•
Ensure that the test box reference microphone is secure. Unplug the white test box
reference microphone. Plug the microphone back in again, ensuring that it ‘clicks’ into
place. Try the calibration procedure again.
•
Unplug the test box reference microphone and plug the On-ear probe microphone in its
place. Try test box calibration using the reference microphone of the On-ear probe
microphone in place of the Test box reference microphone. If this is successful, the On-
ear reference microphone is defective and must be replaced.
Invalid test box calibration
Invalid test box Calibration!
This message may be displayed during the test box calibration procedure. It indicates that
the calibration curve is significantly different from what is expected. It may be accompanied
by a missing test box reference microphone error.
•
Check that the there is no coupler on the coupler microphone and that the
microphones are correctly positioned for test box calibration.
•
Check that the coupler microphone is securely plugged into the correct socket.
•
Follow the suggestions in No test box reference mic. detected.
No on-ear ref. mic. detected
No On-ear reference mic. detected.
Check connection on the On-ear probe microphone.
•
Check that a probe microphone assembly is securely plugged into the PROBE socket to
the left of the test chamber.
•
Unplug the Test box reference microphone in the test chamber and plug the On-ear
probe microphone in its place. Try Test box calibration using the reference microphone
of the On-ear probe microphone in place of the Test box reference microphone. If this
is fails, the On-ear probe microphone is defective and must be replaced.
Invalid on-ear calibration
Invalid On-Ear Calibration!
This message may be displayed during the on-ear calibration procedure. It indicates that
95
Summary of Contents for RM500SL
Page 1: ...RM500SL User s Guide Version 3 4 March 2010...
Page 9: ...RM500SL User s Guide Version 3 4 June 2011 EC Declaration of Conformity 9...
Page 10: ...RM500SL User s Guide Version 3 4 June 2011 10...
Page 37: ...RM500SL User s Guide Version 3 4 June 2011 37...
Page 43: ...RM500SL User s Guide Version 3 4 June 2011 telecoil test 43...
Page 51: ...RM500SL User s Guide Version 3 4 June 2011 51...
Page 57: ...RM500SL User s Guide Version 3 4 June 2011 57...
Page 64: ...RM500SL User s Guide Version 3 4 June 2011 64...
Page 78: ...RM500SL User s Guide Version 3 4 June 2011 Screen tour unaided screen 78...
Page 97: ...RM500SL User s Guide Version 3 4 June 2011 97...
Page 98: ...RM500SL User s Guide Version 3 4 June 2011 98...
Page 99: ...RM500SL User s Guide Version 3 4 June 2011 99...
Page 102: ...RM500SL User s Guide Version 3 4 June 2011 102...
Page 112: ...RM500SL User s Guide Version 3 4 June 2011 112...
Page 113: ...RM500SL User s Guide Version 3 4 June 2011 113...