Verifit
®
User's Guide Version 4.8
©
November 2016
select.
4. Click
to signal average and store a complete passage. Repeat as necessary.
5. Click
to stop the test without recording any test data.
Upon completion of a test recording, a colored Aided Audibar shows the SII for the aided average speech signal.
The difference between the unaided and aided SII score (and the associated length of their respective 'Audibars')
visually quantifies the speech intelligibility improvement for soft speech likely provided by the hearing
instrument being fit. To further demonstrate the improvement, the unaided speech 'banana' may be displayed by
clicking
beside the Unaided Audibar.
Adjusting the Maximum Output Level
See Speechmap screen choices for setup and audiometric data entry. See Using Speechmap.
Follow the instructions in Speechmap screen choices to enter audiometric data and select the fitting rule.
1. Click
in the
3
area to start the third test. Once the test is started, click the [Stimulus] list button to select
the stimulus type as MPO. When you start a new test, the speech envelope of the previous test may be
automatically hidden. See Speechmap Setup to change this behaviour. To change the default stimulus type
and level see Saving test setup.
2. In Test box mode, while the 90 dB sweep is being presented, adjust the hearing instrument's high-level gain
and/or output limiting (e.g. MPO, PC) so that the curve matches the + REAR90 targets (DSL) or approaches
59
Summary of Contents for verifit
Page 1: ...Audioscan Verifit User s Guide 4 8 November 2016...
Page 17: ...Verifit User s Guide Version 4 8 November 2016 EC Declaration of Conformity 17...
Page 92: ...Verifit User s Guide Version 4 8 November 2016 92...
Page 94: ...Verifit User s Guide Version 4 8 November 2016 94...
Page 107: ...Verifit User s Guide Version 4 8 November 2016 Telecoil test results 107...
Page 123: ...Verifit User s Guide Version 4 8 November 2016 123...
Page 148: ...Verifit User s Guide Version 4 8 November 2016 148...
Page 149: ...Verifit User s Guide Version 4 8 November 2016 149...
Page 159: ...Verifit User s Guide Version 4 8 November 2016 159...