12
1:2 Instructions for service and maintenance
Tighten drive motors
It is IMPORTANT to tighten all drive motor nuts at every
service situation. If they are not properly tightened the
fastening might be damaged.
See method 2:4:1, pic. 1- 3 for removing the wheel.
Tighten the three nuts on both sides of the wheelchair
as shown in picture.
WARNING!
Do not use too much force! These are alumin-
ium threads and shall not be over tightened.
Rather use some lock fl uid, such as Loctite.
Instructions for the 6 months service
Motor release
If the motor release is not working as it should, see
method 2:4:4
Change of transversal release shaft.
Check if any of the joints or bars are broken.
Control unit arm lock/unlock in the position chosen
If the joystick arm doesent lock in the position chosen, the cause is
most probably a defect gas spring or release lever.
Se
e method 2:4:12
Joystick arm release, change incl. adjustment
and change of gas spring
Rubber housing for seat lift pillars
It is IMPORTANT to check that the rubber housings are
not defect.
Defect rubber housings can expose the Ketter-pillars and the seat
height system to water and rust.
See methods 2:4:6, pic. 1- 4, Carony Go
2:4:7, pic. 1- 3, Carony Drive
Summary of Contents for Carony Drive
Page 2: ......
Page 4: ...4 ...
Page 6: ...6 ...
Page 10: ...10 0 1 Technical data ...
Page 22: ...22 2 2 Exploded views and components 2 Carony Go foot rest 3 4 2 5 6 7 1 3 4 5 2 6 7 1 ...
Page 26: ...26 2 2 2 Exploded views and components Carony Go Drive lever system 1 2 4 9 7 5 6 8 10 11 3 ...
Page 34: ...34 2 2 Exploded views and components Carony Go Drive seat and armrests 1 8 7 6 3 5 2 10 9 4 ...
Page 62: ...62 22 ...
Page 86: ...86 Notes ...
Page 87: ...87 Notes ...
Page 88: ...Notes ...