Error description
Possible causes
Troubleshooting
covered.
The connection between
blackbody and control box is
defective.
The preset value on the control
box has changed.
foreign objects.
Restore the
connection.
Check the settings
using the supplied
calibration certificate
and correct them
using the arrow keys
on the controller unit.
The user display remains dark. The unit has not been switched
on properly.
The power supply cable has not
been plugged in and the battery
is empty.
Turn on the tablet by
pressing the button on
the side of the screen
(keep pressing until
you get a haptic
feedback).
Establish the
connection between
the tablet and the
embedded
visualization PC..
The user display cannot show
the web page.
The WLAN connection is
disturbed.
Make sure that the
embedded
visualization PC is
switched on and you
are logged in.
Disposal
Like most mechanical and electronic products, this system, of the type
FebriScan, is subject to the regulations for environmentally friendly disposal.
Please contact the manufacturer if necessary!