Bulb and Rod Replacement
Note: Use of any bulb other than that expressly and directly purchased from an
Advanced Ultra-Violet Systems fully authorized source will cause failure in the
disinfection cycle. This may result in electrical failure that is not covered under
warranty and will cause premature device failure. Use of any bulbs other than those
provided by the manufacturer is prohibited. In order to replace a bulb or a quartz
rod, follow these instructions:
Note: UV Box Models that do not have a side access pannel must be returned to the
factory for repair or replacment.
Replacing a Quartz Rod
•
First remove the access panel on the left side of the UV Box. (see figure A)
•
Carefully slide out the broken Quartz Rod. Be cautious of sharp edges.
•
Discard broken rod, making certain to remove all fragments from the inside of the AUVS
Germicidal Enclosure. See:
Cleaning the AUVS Germicidal Enclosure.
•
Gently insert the new Quartz Rod in the vacant slot and replace the access panel.
Replacing UVC Bulbs
•
First remove the access panel on the left side of the UV Box. (see figure A)
•
Carefully slide out and set aside the Quartz Rods to gain access to the UVC Bulb enclosure.
Remove UVC Bulbs. See:
Procedures for Disposal of Bulbs and Special Handling
Information for Broken Bulbs.
•
Replace inoperative UVC bulbs with new AUVS Approved UVC Bulbs only.
Please note: For
maximum efficency, it is recommended that you replace both UVC bulbs, even is one is still
operative.
•
Gently reinsert the Quartz Rods into their slots and replace the access panel.
Figure A