Avidity Science, LLC
Geno Range RO/DI Water System, Series 5 Operation Manual
Routine Maintenance
11
Routine Maintenance
Below is a schedule of routing maintenance intervals required to keep the Geno Water System operating.
Replace Pre-Treatment Module
The pre-treatment module should be replaced every 3 to 6 months depending on the quality of the supply
water. The module is located behind the front panel of the RO machine. You will need
Pre-Treatment
Module TC001
.
Follow this procedure.
1.
Lift the left front panel and remove.
2.
On ,
select
.
3.
Enter the User passcode
1111
and touch
4.
On ,
touch
TC001 Change
and touch
.
Table 3. Schedule of Routine Maintenance
Maintenance Required
Frequency
Preventive maintenance by an Authorized Technician
6 months
Replace Pre-Treatment Module
6 months
a
a. Recommended frequency depends on usage and applies to the stated model only, without modification.
Replace Endure Purification Pack 2
6 months
Replace RO membrane
On recommendation from an Authorized Technician
Replace Air Filter on Storage Tank
12 months
Sanitise RO membrane (RO Clean 1)
12 months or as needed based on bacterial growth.