Clinical accuracy validation method 

 
NT16B is an adjusted mode clinical thermometer 
the validated information for clinical accuracy in each 
adjusted mode are: 
 

GroupA1:   

cb=-0.01°C, L

A

=0.18, σ

r

=±0.08°C 

GroupA2:   

cb= 0.06°C, L

A

=0.22, σ

r

=±0.08°C 

GroupB :   

cb=-0.01°C, L

A

=0.20, σ

r

=±0.07°C 

GroupC :   

cb=-0.01°C, L

A

=0.18, σ

r

=±0.07°C 

 

cb: CLINICAL BIAS 
L

A

: LlMITS OF AGREEMENT 

σ

r:

 

CLINICAL

 

REPEATABILITY

 

 

 

 

Cleaning and Disinfecting 

 

For home use device disinfection, 70% Ethanol or Isopropyl 

alcohol(available in the pharmacy) can be used. 

 

-  Clean the measuring sensor after each use. Use a clean 

cloth or cotton bud that can be moistened with 70 % alcohol. 

-  To clean the entire device, please use a soft cloth slightly 

moistened with a mild soapy solution. Under no 

circumstances may liquid enter the device. Do not use the 

device again until it is completely dry. 

 

   

 

11 

 

 

Applied Standards 

 

This product conforms to the provisions of the EC directive 

MDD(93/ 42/ EEC). The following standards apply to design 

and/or manufacture of the products: 

 

‧  ISO 80601-2-56 

Medical electrical equipment -- Part 2-56: Particular 

requirements for basic safety and essential performance 

of clinical thermometers for body temperature 

measurement 

‧  IEC/EN 60601-1 

Medical electrical equipment- Part 1: General requirement 

for safety 

‧  IEC/EN 60601-1-2 

Medical electrical equipment- Part 2: Collateral standard: 

Electromagnetic compatibility - Requirements and tests 

 

 

FCC Important Information 

 

 

 

a.       FCC Caution: To assure continued compliance, 
(example 

  use only shielded interface cables when 

connecting to computer or peripheral devices). Any 
changes or modifications not expressly approved by the 
party responsible for compliance could void the user

authority to operate this equipment. 
b.       This device complies with Part 15 of the FCC 
Rules. Operation is subject to the following two 
conditions: (1) this device may not cause harmful 
interference, and (2) this device must accept any 
interference received, including interference that may 
cause undesired operation. 
c.       This device complies with FCC radiation 
exposure limits set forth for an uncontrolled 
environment. End users must follow the specific 
operating instructions for satisfying RF exposure 
compliance. This transmitter must not be colocated or 
operating in conjunction with any other antenna or 
transmitter.

 

12 

 

 

Summary of Contents for NT16B

Page 1: ...attery Installation 5 Switching Between Fahrenheit or Celsius 6 Switching between 4 Kinds of Scan Mode 6 Tips for Measuring Human Temperature 7 Measuring Human Temperature 7 Measuring Object Liquid Temperature 9 Memory Function 10 Cleaning and Disinfecting 11 Applied Standards 12 FCC Important Information 12 Error Codes 13 Operating the Bluetooth function 14 Technical Specification 15 EMC Tables 1...

Page 2: ...rmometer 2 Be sure to clean the thermometer lens each time after usage 3 Avoid direct finger contact with the lens 4 No modification of this equipment is allowed 5 It is recommended that user may take 3 temperatures If they are different use the highest reading 6 Do not expose the thermometer to extreme temperature very high humidity or direct sunlight 7 Avoid extreme shock or dropping the device ...

Page 3: ...Product Identification 3 Description of LCD Display 4 ...

Page 4: ...s must be disposed of in accordance with local environmental and institutional policies 5 Dispose of used batteries in accordance with the applicable legal regulations Never dispose of batteries in the normal household waste 5 Switching Between Fahrenheit or Celsius Your thermometer can display results in either degrees Celsius C or degrees Fahrenheit F To switch between Celsius and Fahrenheit whi...

Page 5: ...ymbol will appear on the display If the patient s skin is covered with hair sweat or dirt clean the area and wait 10 minutes before taking a measurement Ensure that the thermometer is held firmly during measurement and that the patient does not move until the measurement is complete Movement can impact the measurement 7 Measuring Human Temperature 3 Position the thermometer under 3cm around 1 inch...

Page 6: ...memory to share with your physician or trained healthcare professional 1 When the device is on press once briefly on the MEM button then pass it again to show the last measurement accompanied by symbol 2 The symbol or symbol will appear with each measurement stored in memory to indicate whether a person or object temperature was taken 3 Each press of the same button recalls a previous measurement ...

Page 7: ...rt 2 56 Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement IEC EN 60601 1 Medical electrical equipment Part 1 General requirement for safety IEC EN 60601 1 2 Medical electrical equipment Part 2 Collateral standard Electromagnetic compatibility Requirements and tests FCC Important Information a FCC Caution To assure continued...

Page 8: ...perating the Bluetooth function What You Need NT16BB Bluetooth IR Ear Thermometer An Android device with Android version 4 3 or above and hardware support for Bluetooth 4 0 An iOS device with iOS version 5 or above and hardware support for Bluetooth 4 0 All devices Apple released since the iPhone 4S including the 4S do the older ones don t Note Please refer to the instruction manual of your smart ...

Page 9: ...equency Interference EMC RFI Readings may be affected if the unit is operated within a radio frequency electromagnetic fi eld strength of approximately 3 volts per meter but the performance of the instrument will not be permanently affected Care should be taken to keep the NT16B at least 6 inches away from R C transmitters to avoid radio frequency interference 2 Avoid keeping the NT16B too close t...

Page 10: ...hermometer is intended for use in the electromagnetic environment specified below The customer or the user of the NT16B TOUCH FREE Infrared Thermometer should assure that it is used in such an environment Immunity test IEC 60601 Test level Compliance Level Electromagnetic environment guidance Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3Vrms 150kHz to 80 MHz 3V m 80MHz to 2 5 GHz 3 Vrms 3...

Page 11: ...r W 150kHz to 80MHz d 1 2 P 80kHz to 800MHz d 1 2 P 800kHz to 2 5GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 19 Explanation of Symbols The CE marking with the Registration Number of the Notified Body This denotes the compliance of European Medical Device Directive 93 42 EEC Follow instructions for use Disposal information Should you wish to dispose ...

Page 12: ...72 N16BMN AV1 2017 03 27 ...

Reviews: