16
Error Codes
When a malfunction or incorrect temperature measurement occurs, an
error message will appear as described below.
Display
Cause
Solution
The temperature measured is
higher than measurement range.
Human mode:
43°C (109.4°F)
Object mode:
100°C (212.0°F)
Operate the thermometer
only between the specified
temperature ranges. If
necessary, clean the sensor
tip. In the event of a repeated
error message, contact your
retailer or Customer Services.
The temperature measured is
lower than measurement range.
Human mode: 34°C (93.2°F)
Object mode:
0°C (32.0°F)
The operating temperature is not
in the range:
15°C~40°C (59°F~104°F)
Operate the thermometer
only between the specified
temperature ranges.