12
requested from supplier using the
contact information provided in this
manual. However, it is advisable to
keep the electromechanical aerosol
equipment at an adequate distance
of, at least, 0.5 m from mobile phones
or other RF communications
transmitters to minimise possible
interference.
RATED power
frequency
magnetic fields.
IEC 61000-4-8 30 A/m c)
50 Hz or 60 Hz
a) The equipment is suitable for use in Home Health Environments and
Professional Health Care Environments limited to patient rooms and
respiratory treatment facilities in hospital or clinics. The more restrictive
IMMUNITY acceptance limits have been considered and applied.
b) Before modulation is applied.
c) This test level assumes a minimum distance of at least 15 cm between
the ME EQUIPMENT or ME SYSTEM and sources of power frequency
magnetic fields.