Configure user data
The
Config
command is used to change the user data of the system. The parameter of this command is
the list of the user data.
Table 10
: Configure user data command with description of the bytes.
Byte
Possible HEX
Description
1 – 4
7B 43 50 00
Start of packet with size (80 bytes)
5 – 6
00 EA
Additional command
7 – 10
00 00 40 40
Concentration gas 1 Offset
(here 3)
11 – 14
CD CC 8C 3F
Concentration gas 1 Span
(here 1.1)
15 – 18
00 00 00 00
Reserved variable for future use (dummy)
19 – 22
00 00 00 00
Reserved variable for future use (dummy)
23 – 26
00 00 00 00
Analog output: Min concentration
(here: 0)
27 – 30
00 00 C8 42
Analog output: Max concentration
(here: 100)
31 – 34
00 00 00 00
Analog output: Offset
(here: -1)
35 – 38
0A D7 23 3C
Analog output: Span
(here 0.01)
41 – 42
00 00 20 41
Integration time
(here: 10)
43 – 46
00 80 BB 44
Concentration gas 2 Offset
(here 1500)
47 – 50
33 33 73 3F
Concentration gas 2 Span
(here 0.95)
51 – 54
00 00 00 00
Diagnostic Mode (1 = active / 0 = inactive)
(here 0)
For further details about Diagnostic mode, contact Axetris AG.
55 – 58
00 00 00 00
Laboratory Mode (1 = active / 0 = inactive)
(here 0)
For further details about Laboratory mode, contact Axetris AG.
59 – 62
00 00 00 40
Analog switch
(here 2)
63 – 66
00 00 00 00
Reserved variable for future use (dummy)
67 – 70
00 00 00 00
Reserved variable for future use (dummy)
71 – 74
00 00 00 00
Reserved variable for future use (dummy)
75 – 78
00 00 00 00
Trailing zeros
79 – 80
A2
7D
End of packet with checksum
As confirmation, the sensor responds with a
Config
-packet without any data.
Please note that this general description may be subject to changes depending on single
product execution.
The device is configured that it always switches from Laboratory mode back to normal after
the device was powered off.
Not all LGD F200P2 module configurations do offer Laboratory mode option
Summary of Contents for LGD F200P2
Page 1: ...Operating Integration Instructions Laser Gas Detection OEM Module LGD F200P2 Axetris AG...
Page 57: ...Axetris AG LGD F200P2 Operating Instructions Rev D 19 09 2018 57 60 Notes...
Page 58: ...Axetris AG LGD F200P2 Operating Instructions Rev D 19 09 2018 58 60 Notes...
Page 59: ...Axetris AG LGD F200P2 Operating Instructions Rev D 19 09 2018 59 60 Notes...
Page 60: ...Axetris AG LGD F200P2 Operating Instructions Rev D 19 09 2018 60 60 Notes...