8
●
When rotary switch being in the positions of mV, logic frequency, diode, resistance and
capacitance, the maximum overload protection voltage will be 250V(effective value),while in the
positions of µA/mA the protection current being 0.64A,and in the position of A.
●
Over range indication OL.
●
When the total voltage of batteries being lower than 6.8V, the symbol for battery display will be
lighted.
●
Fuse being 0.63A/600V(µA/mA end) .
●
Six 7# AAA batteries X 6.
●
Operating temperature:
0
℃
~
30
℃
(relative humidity 0
~
80%)
31
℃
~
51
℃
(relative humidity 0
~
50%)
●
Storage temperature:
-20
℃
~
60
℃
(relative humidity <= 80%)
●
Altitude:
operation
less than 2000m,
Storage
less than 10000m
●
Volume
:
200mm X 100mm X 40mm
●
Weight
:
560g
Range and Accuracy
The below-listed accuracies under different ranges refer to those which are guaranteed by the
meter within one-year calibration, with normal use under the operating temperature of 18
℃
-28
℃
and relative humidity less than 80%.The presentation for accuracy is: ± (**% reading number
of lower digits)
•
AC Voltage
Range
Resolution
Accuracy
40Hz~400Hz
500mV
0.1mV
±(0.8% +4)
5V
1mV
±(0.8% +4)
50V
10mV
±(0.8% +4)
500V
0.1V
±(0.8% +4)
1000V
1V
±(1.0% +4)
Notes: above accuracies can be guaranteed within 10%-100% of the full range.
•
DC Voltage
Range
Resolution
Accuracy
500mV
0. 1mV
±(0.1% +2)