INSTRUCTION MANUAL 

 

_____________________________________________________________________________________________________________________

 

 

 

8.  Balance preparation to work  

 

1.

 

Unpack the balance. 

2.

 

Built-in the scale into horizontal line sector, in place not subjected to mechanical vibrations and strong 
air movements. 

3.

 

Using M12 nut set the line sector precisely in axis of existing line with 3

1

mm intervals from both sides. 

4.

 

Connect  the  ground  wire  to  scale  (label  on  the  scale  indicates  where  it  should  be  connected)  and 
ground the rest of existing line.  

 

 
 
 
 

5.

 

Mount read-out module in convenient place. 

 

6. Connect the supply cable plug to socket with protective contact, when balance is unloaded. 

 

 

 
 

Summary of Contents for 4BA/T Series

Page 1: ...USER MANUAL LINE SCALES 4BA T SERIES FILE 2016 01 19 4BAT 029 bE0106 GB ...

Page 2: ...__________________________________ 18 15 Scale menu _____________________________________________________________________ 19 16 Menu navigation rules_____________________________________________________________ 20 17 Scale setup SEtUP ______________________________________________________________ 25 17 1 Scale calibration CALIb __________________________________________________________ 27 17 2 Aut...

Page 3: ...munication channel choice function rF CHn _________________________________ 55 18 13 Charging accumulators function bAttErY option ______________________________________ 56 18 14 Automatic switching off the scale function AutoOFF ___________________________________ 57 18 15 Statistical calculations function StAt _______________________________________________ 58 18 16 Paperweight calculation PAP _...

Page 4: ...ed for special applications mentioned in Ministry of Labour and Social Policy decree from 11 12 2003 trade tariffs pharmacy recipes medical and pharmaceutical analysis packing of goods For other applications it is recommended to replace verification with calibration Verified balances have the following verification features protection stamps located on balance meter and converters connection box i...

Page 5: ...ication unit e 100g 200g Minimal load Min 2kg 4kg Tare range 300kg 600kg Accuracy class III Work temperature 10 40 C Weighing time 3s Line lenght 0 5m Line diameter 48mm Display cable length 4m Supply 230V 50Hz 8VA Protection step meter IP65 sensor IP68 3 Completation Basic set consists of 1 Scale sensor with line 2 Scale meter 3 Scale suspension M12 screws 4 pcs M12 nuts 16pcs straps 2pcs 4 RS232...

Page 6: ...ta For the balance supply use the socket with protective contact Do not use balance when the cover is removed Do not use balance in explosive atmosphere Do not use balance in locations with high humidity when the cover does not have special protections NAN types In case of damage suspicion switch the balance off and do not use it until it is tested in professional service company 5 Principles of u...

Page 7: ...________________________________ 6 Scale external outputs Separate ground connection scales in stainless steel version must be connected using additional conductor Extensometers are connected permanently RS232C Extensometer Power unit 12V Feeder USB Ground RxD receive ground housing RS232C 1 2 3 4 5 6 7 8 9 ...

Page 8: ... clock is installed and on and nett weight During transmission LabEL communicate is displayed The way of sending data and transmission parameters is set using SErIAL special function Set of send data is set using special function PrInt The following data can be send Header scale type Max d e serial number Operator identification number Successive printout number measurement Identification number o...

Page 9: ...03h command For example Using a program to test RS232 interface program is available in www axis pl programy komputerowe for scale number 1 please write 0201 to log in then SI and write 03 to close communication Asking about scale presence in system testing scale connection with computer Computer Scale S J CR LF 53h 4Ah 0Dh 0Ah Scale Computer M J CR LF 4Dh 4Ah 0Dh 0Ah Displaying a inscription on s...

Page 10: ...30h 2Eh 30h 0Dh 0Ah in order to set low threshold 100kg in scale B150 d 50g the following order should be sent S L 1 0 0 0 0 CR LF 53h 4Ch 31h 30h 30h 2Eh 30h 30h 0Dh 0Ah Setting threshold 2 value option Computer Scale S H D1 DN CR LF 53h 48h D1 DN 0Dh 0Ah D1 DN threshold value maximum 8 characters Scale Computer without response Setting threshold 3 value option Komputer Waga S M D1 DN CR LF 53h 4...

Page 11: ... signs year month day masa 10 signs scale indication mass unit P1 Steering instruction Attention 1 Except variable signs constant signs can also be inscribed e g factory name product name and so on 2 In standard only one label pattern is possible to printout number 0001 Using bigger amount of patterns other label numbers is possible thanks to LAbEL special function 3 To achieve label printout labe...

Page 12: ... place not subjected to mechanical vibrations and strong air movements 3 Using M12 nut set the line sector precisely in axis of existing line with 3 1mm intervals from both sides 4 Connect the ground wire to scale label on the scale indicates where it should be connected and ground the rest of existing line 5 Mount read out module in convenient place 6 Connect the supply cable plug to socket with ...

Page 13: ...performed by pressing T key 3 Weighed mass should be placed in the platform centre Do not drop weighed objects on the platform Do not overload the balance over 20 of maximum load 4 The weighing result should be read during the indicator lighting which indicates the result stabilization 5 When there is no weighing but the balance should be ready for operation it may be switched off by I key It caus...

Page 14: ...pany to perform balance adjustment 11 Balance adjustment Balance adjustment must be performed by authorized service company only as it is connected with necessity of seals breaching required for warranty purposes 12 Meter keys and indicators LCD display version Keys I switch on switch off standby T tare menu position confirmation next digit when inscribing 0 zeroing scrolling menu increasing digit...

Page 15: ...ARE and key AUT autotare function on T tare memory function on TOTAL FILTER SUM special functions indicators ct n g m2 lb mg unit indicators pcs pieces counting indicator n measurements quantity indicator total function OFF scale turned off standby MIN weighing result under threshold I thr function OK weighing result between threshold I and II MAX weighing result above threshold II BAT battery dis...

Page 16: ...dicator unloaded scale result stabilization indicator NET net weight after using T key W1 first range on in two range scale W2 second range on in two range scale HOLD indication locked concerns LOC and UP functions FUNC special function turned on PCS pieces counting g kg lb unit weight unit g gram kg kilogram lb pound unit other MIN weighing result under threshold I thr function OK weighing result...

Page 17: ..._______________________ 13 Start up Leave the pan empty plug a scale to the mains with a ground contact The scale proceeds with following start up actions Display test Meter type displaying Program version The scale is now ready to work Attention Displaying program version means positive result of all tests 0 0 kg kg ...

Page 18: ...y Zero indication and 0 indicator mean that the scale is ready to work After putting container package tare the scale using T key NET indicator will show up Put on weighted object and readout net weight NET indicator shows that scale indicates net weight In order to readout gross weight press key B G indicator shows that scale indicates gross weight Press again key in order to come back to net ind...

Page 19: ...In out of the box scale after pressing MENU key only SEtuP option it contains all configuration options is available One of the configuration options is Menu that is used to create personalized menu To add a function to personalized menu press T key when the function is indicating Chosen function is indicated with o sign on the left side of display After adding all necessary functions press out in...

Page 20: ...nds and then the scale sequentially displays next menu positions Choosing menu position option is done by pressing T key when it is displayed on the screen After choosing position option usually several options show up on turning on selected option OFF turning off out out to menu Accelerated working with menu Menu first position is displayed for about 7s During this time user can view next positio...

Page 21: ...ey working method During standard weighing key is used to switch between net and gross indication When special function e g PCS is turned on using key enables to go back to standard weighing mode Sign o on the left side signalizes that special function is turned on and user can go back to function mode by pressing key 0 0 0 kg 0 0 0 kg kg kg ...

Page 22: ... normal weighing temporary 5s readout resolution increase is possible Return to normal indication is made automatically Inscribing numerical values Inscribing numerical values is needed in some special functions e g tArE function requires to inscribe tare values Keys 0 increasing digit value decimal point T next digit position MENU end of inscribing 0 0 kg kg 5s 0 kg ...

Page 23: ...ation exit autozeroing autozeroing on autozeroing off lasts 10min exit settings creating menu added to menu choosing product choosing user pieces counting function actual unit selection percentage conversion function animals weighing function tare memory bank maximum value function force and torque value function summation series of measurements function threshold values comparing function statist...

Page 24: ...ansmission type Stab no StAb Auto Cont rEMoVE protocol LonG EPL EPL_A EPL d PEn 01 cooperation with barcode readers MJ 4209 port 2 as above printout data configuration transmission port 1 header scale type Max d e serial nr scale operator id number user name succesive printout measurement number product identification number product barcode number product name pieces quantity PCS function product ...

Page 25: ...____________________________ bAttEry AUto OFF ZEro dEFAULt SErVICE out turn on off accumulator charging if the scale is equipped with accumulator automatic turning off saving accumulator power as above scale start zero inscribing factory zero restore default settings for all options options only for service exit ...

Page 26: ...d user menu CALIb scale sensitivity calibration AutoZEro ing self maintaining zero indication unloaded scale UnIt weight unit selection SErIAL setting serial ports Print transmission printout data selection FILtEr anti disturbance filter b_LIGHt backlight setting Ad420 analogue out configuration FIrMW are updating software only for service dEFAULt reset to factory settings sample of using in chapt...

Page 27: ...CAL StP calibration with external weight confirmation of succesive steps MENU key out leave without changes Press T key when CAL StP option appears calibration in two steps Press T key when weight value used for calibration is indicating or use othEr option and inscribe proper value keys 0 T Press MENU and wait for writing zero to the scale When LOAD message appears put standard of mass on the pan...

Page 28: ...able zero indication if the pan is empty or if zero indication was acquired by pressing T key To turn on the function use MENU key and using T key choose AutoZEr and then Aut on To leave the function press MENU key then with T key chose AutoZEr and Aut OFF Note 1 AUt sign occurs only in scales with LCD display 2 In scales with 0 key active function changes name into AutoZE autozeroing and works on...

Page 29: ...1mg 0 001g milligram KGrAM 1kg 1000g kilogram Pound 1 lb 453 592374g English pound OunCE 1oz 28 349523g ounce OunCEt 1ozt 31 1034763g pharmaceutical ounce GrAIn 1gr 0 06479891g grain PennYW 1dwt 1 55517384g jewellery mass unit GrAM 1g gram The way of choosing carats as weighing unit is shown on the example Attention In scale with LED display designations of mass units lb kg oz ozt ct are not displ...

Page 30: ...s baud rate bAud 4800 9600 115 200bps number of bits in single char bitS 7 8 parity control PArItY nonE no control Odd nonparity Even parity control scale number in network nr if the scale doesn t work in network the number must be 0 transmission through serial interface SendInG StAb transmission after key is used and result is stable noStAb transmission after key is pressed without need of stabil...

Page 31: ... choose this option to zero counter Prod Id product number Prod bA product barcode inscribed or scanned Prod nA product name Count counting result PCS function APW unitary mass PCS function netto net mass tArE current tare value GroSS gross mass totAL total mass totAL function Attention If Prod Id or USEr Id is chosen it is possible to inscribe quickly their new values with omission of main menu I...

Page 32: ...intout during normal weighing with clock option all printout positions deactivated Sample printout during normal weighing some printout positions activated 20 07 kg 20 04 kg 20 04 kg 20 07 kg 2012 11 08 10 01 20 04 kg 2012 11 08 10 01 20 04 kg 2012 11 08 10 01 BA30 MAX 30kg e d 0 01kg S N ID OPER 000001 DATE 2012 11 08 TIME 12 26 NO 3 ID PROD 01 COUNT 0 PCS APW 0 000 g NET 3 08 kg TARE 0 00 kg GRO...

Page 33: ...g the work mode of scale display backlight b_L OFF switch backlight off b_L on switch backlight permanently on b_L ECO switch off after 30 seconds of inactivity no load changes and no key operation b_L bAt like above but when powering from accumulators only out out without changes Switching backlight off causes decrease of energy consumption by the scale what is important during powering from accu...

Page 34: ...__________________________ 17 7 Analog out configuration AnALoG This option enables to set up analog out 4 20mA or 0 10V working method used e g in PLC regulators AnG rnG inscribing Max value AnG CFG working mode configuration PLUS workmode for only positive values MinuS only for negative values both for both 0 kg ...

Page 35: ...ess MENU key When ZErO is displayed press T key On the display a sign ZEr Cod will show up momentary and the a dash on last digit position To enter code in new scale 1234 use keys 0 increasing digit T next digit MENU end of inscribing The following options appear successively on display ZEr Cod enter new secure code value ZEr SEt enter new zero value Using T key choose ZEr SEt Direct result from A...

Page 36: ...on tArE maximum value indication function UP force measuring function nEWton statistical calculations StAt paperweight calculation function PAPEr and functions that require additional equipment to be completely functional option with accumulator supply Setting accumulators charging bAttErY Automatic switching off scale function AutoOFF options with the clock setting current date and time function ...

Page 37: ... datasheet form where each product has one row and each column have product data This way created database saved in csv extension with semicolons can be send to scale using Scale Database software and scale s serial interface Scale Database is available on our webpage www axis pl en Database and possibility to cooperate with external devices printer label printer barcode reader and computer enable...

Page 38: ...s to send it to the port to printer computer together with actual weighing result LIMIt searching through products from database if product is found then Found communicate appears and all product data is recalled if there is no product in database then not Found communicate appears Prn_P sending all products database to port To inscribe data use EdIt option and keys 0 increasing digit T next dixit...

Page 39: ...dicates one of communicates SCAn barcode from outside the base accepted Std mode not Found barcode from outside the base not accepted LIMIt mode Found product barcode found in database and data recalled Attention If the scale doesn t indicate any communicate check barcode reader connections port configuration and transmission protocol SErIAL function Other fast way is to press and hold MENU key Pr...

Page 40: ...r To each weighing results transmission a set of product and user identification data is added The set is activated in Print option 17 5 chapter Available data from products and users base Print SEtuP option USEr Id user identification number USEr nA user name inscribed from PC or scanner Prod Id product identification number Prod bA product barcode inscribed or scan Prod nA product name inscribed...

Page 41: ...PCS APW set unitary mass directly PCS rS inserting number of details in a sample and receiving of their mass from other scale connected by RS 232C It is advised that single piece weight is not less than one reading unit and sample weight used in first phase is bigger than 100 reading units To leave function press MENU key and then using T key chose PCS and PCS oFF Note 1 APW too LOW communicate si...

Page 42: ...100 second phase measuring specific sample as a percentage of the reference sample Weighing result is displayed in different format depending on the weight value of reference sample The function has the following options PEr oFF disable the function PEr on set current scale indication as 100 and activate percentage weighing out exit without changing settings Note 1 PEr Err message informs that ref...

Page 43: ... user should be saved in scale memory according to printer manual Label pattern choice is made by inscribing label number using LAbEL function Press MENU button When LAbEL is displayed press T key Actual label number will show To enter new label number press T key to exit function without number change press MENU To inscribe label number use keys 0 digit increase T next digit MENU end After enteri...

Page 44: ...ing the scale LOC Prn the measurement initiated manually by pressing key When LOC on is displayed press T key Tare the scale using T key if necessary and place the animal on the pan Wait until the weighing result is averaged scale display blinks Then scale will show stable averaged result and will send it through serial port The result remains on display for about 30 second Important notes 1 The l...

Page 45: ...tare value to memory After pressing MENU key and choosing tArE function using T key the following options are available tAr 0FF leave the function tAr on activate the function with the previous tare value tAr sample tare value from the pan tAr SEt enter tare value with keys 0 T and MENU out printout a setting value of tare Press T key when tAr SEt is displayed By pressing T key choose proper memor...

Page 46: ...will show up tAr 01 02 10 Cells with inscribed value are marked with o sign on the left side active value marked with ATTENTION In scales with LED display cells with inscribed value are marked with Choose proper memory cell using T key tArE function is activated with chosen tare value Moreover the scale will indicate nett weight weight on the pan minus tare values Using T key or 0 while empty pan ...

Page 47: ...is function allows holding maximum or minimum value that is indicating at the moment Before measurement scale should be tared Function has following options UP oFF function off HIGH holding maximum value LOW holding minimum value Pressing T key will cause result zeroing Note Autozeroing function and the stabilisation indicator are deactivated when UP function is running 0 0 kg kg 1kg ...

Page 48: ...ivation causes displaying result in force units N Press MENU key Using T key choose NEWto function Function has several options nEW oFF function off nEW on measurement in Newtons ArM torque measurement arm length should be inscribed in meters using T 0 and MENU keys Attention Units convertion from mass kg to force N is made for acceleration of gravity g 9 80665m s2 Note 1N 0 1019kg 0 kg 0 ...

Page 49: ...ot working with receipt printout after each measurement tot working without receipt printout tot CFG saving measurement mode using key Manual after taking off the load auto Press T key when tot is displayed Perform measurement series by pressing key for storing results into total register In order to print and display results enter the function by choosing totAL and tot Prn option from menu The re...

Page 50: ...clock Date and Time do not appear on printout Maximum number of measurements is 99 999 Maximum total load 99 999 000d The weighing unit of the total value from the register Total is the same as the weighing unit stated on the keypad or is 1000 times greater what is signalled by o indicator at the left of the display If the registered value is too big to be displayed E communicate appears on the di...

Page 51: ...ignal greater than upper threshold the scale signals MAX red colour long sound signal The checkweighing results can be use to control optical indicator Indication mode batching devices Batching mode Standard scale is set for cooperation with optical indicator On outputs P1 P3 Relays socket short circuit states appear as result of comparison scale indication with threshold values On the chart below...

Page 52: ...g T key The following options for entering thresholds are displayed SEt LO set lower threshold value SEt HI set upper threshold value SEt ZEr set zero signalisation threshold Using T key select SEt LO option Set lower threshold value using the following keys 0 digit increase decimal point T move to next digit MENU finish Then select SEt HI option and enter upper threshold value Choosing SEt ZEr op...

Page 53: ...des e g 1N4148 It is advised to use MS3K P electronic board sold separately consisting of RM96P transmitters with DC24V input voltage and AC250V 3A output Important notes 1 After switching the scale on both thresholds are set to maximum values 2 When setting upper threshold value pay attention that its value is not below lower threshold value 3 Setting lower and upper threshold value is possible a...

Page 54: ...rent indication key dAt SEt change current date and time dAt PIn data and time secure password to prevent from changing date and time by unauthorized personel dAt For data printout in USA or EU format The example at the left presents how to set current date and time using dAt SEt option After setting proper date and time activate it with dAt on option UE rrrr mm dd gg mm USA mm dd rrrr gg mm AM PM...

Page 55: ...he scale and a pilot In scale and in pilot the same radio channels must be chosen Function should be used when communication is disturbed by other devices that use the same communication channel Press MENU key and choose rF CHn by pressing T key The following communicates will appear on display Na wyświetlaczu pojawią się kolejno CHn 01 channel 1 CHn 02 channel 2 CHn 16 channel 16 out out without ...

Page 56: ... during work with feeder and checking their power level The function has the following options bAt OFF charging off option required if ordinary batteries are used bAt on charging on accumulators are being charged even after switching scale off using I key bAt VoL reading power level of accumulators in go back to mass indication pressing MENU key out leave without changes An attempt of charging ord...

Page 57: ... is helpful in scales supplied from accumulators The function causes scale to switch off automatically Press MENU key When AutoOFF is displayed press T key The following options appear successively on display AOF oFF deactivate function AOF on activate function scale turns off after 5 minutes of not making any actions AOF bAt as above but only when supplied from accumulators Out out without change...

Page 58: ...cur after the scale is loaded and its indications stabilize After each loading printout is made with number of measurements result date and time if clock is installed and the function is activated For the obtained measurements series the scale evaluates n number of samples sum x sum of all samples n x x sum _ x average value sum x n min minimal value from n samples max maximal value from n samples...

Page 59: ...bing tolerance in Sta tAr automatic tare on off StA CFG function configuration Auto Automatic work samples are confirmed after loading the scale and indication stabilization ManuAL manual work confirmation is made by pressing key out exit from function Remember first to inscribe nominal weight value and tolerance mentioned above After that push T key when StA o is displayed Put on successive objec...

Page 60: ...ERAGE average weight as Total n MIN minimum weight in n samples MAX maximum weight in n samples ST DEV standard deviation ST DEV standard deviation percentage To finish work with this function and zeroing result register press MENU key and then when StAt and Sta oFF is displayed press T button Statistics function cooperation with computer and Printer Scale can be equipped with two serial ports mar...

Page 61: ... Menu To enter the function press T key when PAPEr is displayed Following options show on the display PAP oFF turn off the function PAP on turn on PAP n inscribing number of paper pieces on pan PAP ArE inscribing surface of single piece in m2 Press T key when PAP n and PAP ArE is displayed Enter number of samples using 0 increasing digit T next digit MENU end of inscribing Press T key when PAP ArE...

Page 62: ...en mechanically damaged 19 It is forbidden to make any repairs by unauthorised persons 20 To repair the scale please contact our nearest service Error communicates Communicate Possible cause Remedy C 1 6 over 1 min selftests failed if displayed more than 1 minute contact an authorised service unLOAd SErvic the scale was switched on with loaded pan remove a load from the pan mechanical damage of th...

Page 63: ...agnetic field immunity test harmonized with the directive 2004 108 WE Electromagnetic compatibility Moreover scales with the following markings on the name plate 16 the number of the Notified Body responsible for EC verification 1443 two digit number of the year of EC verification a green metrology sticker with M mark a protective seal affixed by the Notified Body comply with the requirements on t...

Page 64: ...nt range Max1 50 of maximum load mostly Max2 100 of maximum load and adequate reads digit d1 i d2 d1 d2 Double range options causes change of scale s operation after turn on in small mass range 0 Max1 scale displays result with reading unit d1 when the load pass the Max1 scale changes reading unit on d2 from this moment scale displays result with reading unit d2 on all measure range 0 Max2 return ...

Page 65: ...INSTRUCTION MANUAL 65 _____________________________________________________________________________________________________________________ Notes ...