Reference Section
•
107
Holder Maintenance
Cleaning
For lowest noise, keep the holder clean. Frequently rinse the holder with
distilled water. If more thorough cleaning is required, briefly wash in ethanol
or mild soapy water. Never use methanol or strong solvents.
Replacing the Silver Wire
To replace the silver wire, insert the nonchlorided end through the hole of the
silicone seal and bend the last 1 mm of wire over to an angle of 90°. Press the
wire into the back of the barrel making sure that the silicone seal is flush with
the back of the barrel. Slip the threaded collar over the back of the barrel.
Adapters
HLR-U right-angle adapters allow the HL-U series holder to emerge at 90° from the
headstage. Use the HLR-U with the HL-U holder.
HLB-U BNC-to-Axon adapter allows conventional BNC-type holders to be used
with Axon U-type headstages. Use the HLB-U with all U-type CV and HS
headstages. These headstages have a threaded white Teflon collet.
Input/Output Connections
•
Description of the different connectors on the front and rear panels of the
MultiClamp 700B main unit.
•
See also External Command Inputs, Oscilloscope Triggering, Mode.
Front Panel
Inputs
COMMAND: Voltage or current commands to the MultiClamp 700B are
accepted at this input. The External Command Sensitivity is set in the
Gains panel under the Options toolbar button.
Chapter 5