Reference Section
•
117
•
See also Input/Output Connections.
The signal available at this BNC connector is intended to be used as an external
trigger for an oscilloscope when internal commands (Seal Test), Tuning) or Pulse
are activated, or to indicate the Mode state of the amplifier (commanded either
externally by the Mode BNC or internally by the Auto Mode switch feature). The
Sync Output is a 0 to (approximately) 5 V step aligned with the onset of the Seal
Test, Tuning or Pulse step. Or, when following Mode, 5 V corresponds to VC,
while 0 V corresponds to IC. See the Options / General tab to select the function of
the SYNC output.
Output Zero
•
Subtracts the steady-state current offset (in VC mode) or voltage offset (in IC
mode).
•
Activated by pressing the
button in the Output Zero box, or by
checking the checkbox and manually adjusting the value to the left of the
button.
•
See also Leak Subtraction, Bridge Balance.
The purpose of this control is to zero the output, that is, to remove the DC voltage.
Output Zero works by sampling the current or voltage over a ~70 ms time window
immediately after pressing the button, and then subtracting this value from all
subsequent Primary Output signals. Unlike Leak Subtraction or Bridge Balance, it
does not account for currents or voltages that change as a result of time-varying
command pulses; it simply provides a constant offset adjustment.
The Auto Output Zero only affects the signal on the Primary Output. In other
words, the cell is not affected by the Output Zero command. No other input or
outputs are affected.
Output Zero is useful for recording small signals that are riding on a large, constant
offset current or voltage. However, in general we recommend that it not be used,
since potentially useful information about the biological signal is lost.
Chapter 5