130
•
Reference Section
For R
s
compensation to function properly, whole cell compensation must have been
adjusted and the Whole Cell checkbox must be checked. Whole cell compensation
provides estimates of R
s
and C
m
, which together determine the shape of the
correction current that is injected through capacitor C2 (Figure 5.20). Note that this
C2 correction current does not improve the speed of clamping of the cell; rather, it
charges the membrane capacitance as slowly as before but in a way that is invisible
to the user, because it bypasses the feedback resistor in the headstage.
Figure 5.21.
Schematic whole-cell compensation circuit.
MultiClamp 700B Theory and Operation, Copyright 2005 Axon Instruments / Molecular Devices