Reference Section
•
137
capacitance before the change in membrane resistance occurred. Since this
time constant depends on the parallel value of membrane resistance and the
electrode series resistance, this error could become substantial.
If the cell input resistance becomes comparable to, or less than, the electrode
resistance, the whole-cell patch clamp technique will probably not work. In this
situation it would be preferable to use a discontinuous (chopped) single-electrode
voltage clamp, such as the Axoclamp.
SoftPanel Configuration
The SoftPanel is an optional instrument that provides knob and button control in
place of mouse gliders and clicks in the MultiClamp 700B Commander software.
The SoftPanel is merely a hardware extension of the Commander, and replicates the
many Commander control functions of the MultiClamp 700B.
The SoftPanel comes with a magnetic overlay with pre-defined functions assigned
to the various knobs and buttons. However, the SoftPanel can easily be re-
configured in the MultiClamp 700B Commander software. Click on the Configure
SoftPanel toolbar icon (
) to access the menus for re-configuring each knob or
button.
After assigning the desired functions to each knob or button, remove the pre-
defined magnetic overlay to reveal the erasable surfaces at each knob or button.
Re-label the position with the appropriate function using a marking pen. (Sharpie®
pens are appropriate on this special surface.)
Figure 5.22
Chapter 5