Important Safety Information
•
159
Important Safety Information
DISCLAIMER
THIS EQUIPMENT IS NOT INTENDED TO BE USED AND
SHOULD NOT BE USED IN HUMAN EXPERIMENTATION OR
APPLIED TO HUMANS IN ANY WAY.
WARNING
IF THIS EQUIPMENT IS USED IN A MANNER NOT SPECIFIED
BY THE MANUFACTURER, THE PROTECTION PROVIDED BY
THE EQUIPMENT MAY BE IMPAIRED.
Power-Supply Voltage Selection and Fuse Changing
Supply Voltage
The MultiClamp 700B can be directly connected to all
international supply voltages. The input range is from 100 to
240 V~. No range switching is required. Alternatively, the
instrument can be powered by a DC voltage of 120 to 310 V.
Changing the Fuse
The MultiClamp 700B uses a 250 V~, T2A, 5 x 20 mm fuse.
In the event of fuse failure, disconnect the power cord.
Before changing the fuse investigate the reason for its failure.
To change the fuse:
1. Disconnect the power cord.
2. Use a screwdriver or a similar device to rotate the fuse
holder counterclock-wise.
3. Replace the fuse with another fuse of the same rating.
4. Reconnect the power cord.
Basic Equipment Setup and Safety
1. Connections: Use the included IEC power cord to connect
the instrument to a GROUNDED power receptacle.
2. Mounting: Table or rack.
3. Assembly: The headstage connects to the instrument
through the rear panel, 25 pin D-sub connector marked
"Headstage". Power should always be turned OFF when
connecting headstages to the main unit.
4. Use: Do not operate this equipment with covers or panels
removed.
5. Cleaning: Wipe the headstage connector with a damp cloth
to clean salt spills. Avoid spilling liquids on the headstage.
The Teflon input connector should be kept very clean.
Effective cleaning can be done by swabbing carefully with
denatured alcohol or deionized water. If possible, avoid the
use of Freon since it is thought to be detrimental to the
environment.
Safe Environmental Conditions
1. Indoor use.
2. Mains supply fluctuations: not to exceed ±10% of the
nominal voltage.
3. Temperature: between 5 ºC and 40 ºC.
4. Altitude: up to 2000 m.
5. This instrument is designed to be used under laboratory
conditions. Operate in a clean, dry environment only. Do
not operate in a wet or damp environment.
Static Precautions
If you are in a laboratory where static is high (
i.e.
, you hear
and feel crackles when you touch things), you should touch a
grounded metal object immediately before touching the
headstage.
Shipping the MultiClamp 700B
The MultiClamp 700B is a solidly built instrument designed to
survive shipping around the world. However, in order to avoid
damage during shipping, the MultiClamp 700B must be
properly packaged.
In general, the best way to package the MultiClamp 700B is in
the original factory carton. If this is no longer available, we
recommend that you carefully wrap the MultiClamp 700B in at
least three inches (75 mm) of foam or "bubble-pack" sheeting.
The wrapped MultiClamp 700B should then be placed in a
sturdy cardboard carton. Mark the outside of the box with the
word FRAGILE and an arrow showing which way is up.
We do not recommend using loose foam pellets to protect the
MultiClamp 700B. If the carton is dropped by the shipper,
there is a good chance that the MultiClamp 700B will shift
within the loose pellet packaging and be damaged.
If you need to ship your MultiClamp 700B to another location,
or back to the factory, and you do not have a means to
adequately package it, Axon Instruments can ship the proper
packaging material to you for a small fee. This may seem like
an expense you would like to avoid, but it is inexpensive
compared to the cost of repairing an instrument that has
sustained shipping damage.
It is your responsibility to package the instrument properly
before shipping. If it is not, and it is damaged, the shipper will
not honor your claim for compensation.
Important Safety Information