12
•
Installation and Basic Operation
50 G Range
1. Press the “Options” button, choose the Gains tab, and select the 50 G
feedback resistor in the Voltage Clamp pane. Return to the main
MultiClamp 700B Commander window.
2. Plug the PATCH connector of the PATCH-1U model cell into the
CV-7 headstage.
3. Set Seal Test amplitude to 100 mV and frequency to 50 Hz, then check
the box to make it active.
4. Press Auto Cp Fast to remove the bulk of electrode capacitance
transient.
5. The resulting waveform should be square, except for an initial
overshoot (possibly twice the size of the steady-state response) that
settles to the baseline in about 1 to 2 ms. The rise time to the peak of
the overshoot should be about 50
µ
s. The steady-state amplitude
following the transient should be ~500 mV
p-p
(
±
50 mV).
5 G Range
1. Change the feedback resistor to 5 G in the Options / Gains menu.
2. Press Auto Cp Fast.
3. The step response should be ~50 mV
p-p
(
±
5 mV).
500 M Range
1. Press the “Reset to Program Defaults” button. (By default, the 500 M
range is selected.)
2. Check Seal Test and set the amplitude to 25 mV.
3. Plug the CELL connector of the PATCH-1U model cell into the CV-7
headstage.
4. Press the Auto Whole Cell button.
5. Press Auto Cp Fast button.
6. The step response should be ~25 mV
p-p
.
MultiClamp 700B Theory and Operation, Copyright 2005 Axon Instruments / Molecular Devices