11. Now that the capacitance transients are compensated, it will be possible to
increase the amplitude of the Seal Test pulse without overloading the
MultiClamp 700B. Set the Seal Test amplitude to 100 mV by placing the
cursor over the display (10 mV), double clicking and typing 100 <Enter>.
Clear steps should now be visible on the oscilloscope, with amplitudes of about
5 mV.
Figure 3.14
With the Primary Output: Membrane Current gain set at 0.5 V/nA, this is
equivalent to 10 pA. Hence the resistance of the model patch is calculated from
Ohm’s Law to be R = V/I = 100 mV/10 pA = 10 G
Ω
. Alternatively, check the
Resistance checkbox under the Channel 1 meters.
Tutorial 4 – Whole-Cell Configuration: Voltage Clamp
1. Reset to Program Defaults and set Seal Test frequency to 200 Hz. Plug the
CELL connector on the model cell into the CV-7 headstage.