Site Requirements
The GenePix Professional 4200A is designed to operate optimally if the site requirements are met. As with any
precision optical instrument, care should be taken to maintain a low-dust, low-vibration environment. Temperature
and humidity extremes may compromise performance.
Environmental Temperature: 50–86º F (10–30º C).
Environmental Humidity: 5–95% non-condensing.
Power Requirements: Universal Input 50-60 Hz, 100–240V~AC, 2A max.
Space Requirements: 25.6 in deep × 16.9 in wide × 13.4 in high (65 cm × 43 cm × 34 cm); plus additional 6 in
(15.2 cm) clearance.
Weight Requirements: Sufficient to support 99 lbs (45 kg) with minimal vibration.