18
S
etting up the
High Alarm
and the
Low Alarm
enables to activate
the red LED flash of the Data Logger. (See Dia. F-1)
-Select the channel you need: Air temp./external temp./ Air RH%.
If the logger not yet be set any alarm before, the default will be:
o
-
High Alarm temperature defaults at 85 C.
o
-
Low Alarm temperature defaults at -40 C.
-
High Alarm Relative Humidity defaults at 100%.
-
Low Alarm Relative Humidity defaults at 0%.
If the logger used to have alarm setting before, then, the default will
be the same as your last setting. You could choose to change it or not.
User simply select desired High value or Low value by scrolling up
for increasing or scrolling down for decreasing .
If a temperature or RH% is higher than HI setting or less than LO
setting, ALM LED will flash until you download to a PC.
ALARM
SETTING
Diagram F-1:
Alarm Setting
While in 8829, could not set alarm of temperature and RH at the same.
The "channel" column will be disable in 8829. Please choose the
parameter you want to have alarm from "Unit" column and then set
the value as Diagram F-1
NOTE: