133
Input type
Measurement
range
Reference
range
Indication accuracy
Display
resolution
-200.0 to +250.0 °C
±13.8 mV
-200.0 to 0 °C ±0.3 % FS ± 1 digit
0 to +250.0 °C ±0.15 % FS ± 1 digit
0.1 °C
-200.0 to +500.0 °C
±27.6 mV
-200.0 to 0 °C ±0.3 % FS ± 1 digit
0 to +500.0 °C ±0.15 % FS ± 1 digit
0.1 °C
U
-200.0 to +600.0 °C
±69.0 mV
-200.0 to 0 °C ±0.3 % FS ± 1 digit
0 to +600.0 °C ±0.15 % FS ± 1 digit
0.1 °C
-200.0 to +250.0 °C
±13.8 mV
-200.0 to 0 °C ±0.2 % FS ± 1 digit
0 to +250.0 °C ±0.1 % FS ± 1 digit
0.1 °C
-200.0 to +500.0 °C
±27.6 mV
-200.0 to 0 °C ±0.2 % FS ± 1 digit
0 to +500.0 °C ±0.1 % FS ± 1 digit
0.1 °C
Thermo-
couple
L
-200 to +900 °C
±69.0 mV
-200 to 0 °C
±0.2 % FS ± 1 digit
0 to +900 °C
±0.1 % FS ± 1 digit
1 °C
-140.0 to +150.0 °C
160
Ω
±0.15 % FS ± 1 digit
0.1 °C
-200.0 to +300.0 °C
220
Ω
±0.1 % FS ± 1 digit
0.1 °C
Pt100
-200.0 to +850.0 °C
400
Ω
-200.0 to +700.0 °C
±0.1 % FS ± 1 digit
700.0 to 850.0 °C
±0.15 % FS ± 1 digit
0.1 °C
-140.0 to +150.0 °C
160
Ω
±0.15 % FS ± 1 digit
0.1 °C
-200.0 to +300.0 °C
220
Ω
±0.1 % FS ± 1 digit
0.1 °C
JPt100
-200.0 to +649.0 °C
400
Ω
±0.1 % FS ± 1 digit
0.1 °C
Pt50
-200.0 to +649.0 °C
220
Ω
±0.1 % FS ± 1 digit
0.1 °C
Resistance
temperature
detector
(RTD)
Pt-Co
4.0 to 374.0 K
220
Ω
4.0 to 50.0K
±0.3 % FS ± 1 digit
50.0 to 374.0K ±0.15 % FS ± 1 digit
0.1 K
-13.80 to +13.80 mV
±13.8 mV
-27.60 to +27.60 mV
±27.6 mV
-69.00 to +69.00 mV
±69.0 mV
10 µV
-200.0 to +200.0 mV
±200.0 mV
-500.0 to +500.0 mV
±500.0 mV
100 µV
-2.000 to +2.000 V
±2 V
-5.000 to +5.000 V
±5 V
1 mV
-10.00 to +10.00 V
±10 V
-20.00 to +20.00 V
±20 V
DC voltage
-50.00 to +50.00 V
±50 V
±0.1 % FS ± 1 digit
10 mV
Notes:
•
The indication accuracy applies under standard conditions.
•
The thermocouple input (internal reference junction) does not include reference junction compensation accuracy.
K, E, J, T, R, S, B, N:
U (Cu-CuNi), L (Fe-CuNi):
W-WRe5, WRe5-WRe26, PR40-20,
NiMO-Ni, CR-AuFe, Platinel 2:
Pt100:
JPt100:
Pt50:
IEC584, JIS C1602-1995
DIN43710
ASTM
IEC751 (1995), JIS C1604-1997
JIS C1604-1981, JIS C1606-1989
JIS C1604-1981
Note) Indication equivalent to max. 1 mV sometimes fluctuates under EMC Directive test conditions.