background image

AB-5702 

Specifications 

Basic specifications 

Item Specification 
Rated voltage 

100 V AC to 240 V AC, 50 Hz/60 Hz 

Operating voltage 

85 V AC to 264 V AC, 50 Hz/60 Hz 

Power supply 

Power consumption 

6 VA 

Ambient temperature 

0 °C to 50 °C 

Ambient humidity 

10 %RH to 90 %RH (non-condensing) 

Environmental 
conditions 

Vibration 19.6 

m/s

2

 max. at 16.7 Hz for 0.2 hrs 

Ambient temperature 

-20 °C to 60 °C 

Ambient humidity 

10 %RH to 90 %RH (non-condensing) 

Vibration (transport) 

9.8 m/s

2

 max. at 10 Hz to 150 Hz (in package) 

Transport/storage 
condition 

Vibration (storage) 

3.2 m/s

2

 max. at 10 Hz to 150 Hz 

Installation of location 

Inside of a control panel 

Normal 

Flashing every second 

SC-bus error 

Flashing every 0.25 seconds 

Initializing ON 
Major alarm 

ON 

Minor alarm 

Repetition of 1 second ON and 0.25 second OFF 

LED indications 

Power OFF 

OFF 

Weight 

230 g (without package) 

Material 

Modified PPE resin 

LED: Light-emitting diode 
PPE: Polyphenylene ether 
 

Input/output specifications 

Item I/O 

specification 

Connection

Wire specification 

Status alarm input 

Applied current/voltage: 5 mA DC typ./12 V DC typ. 
Connectable output: Dry contact, transistor output 

M3.5 terminal JIS CVV 1.25 mm

2

 or JIS IV 1.25 mm

2

 

Max. 50 m long 

ON/OFF output 

Output type: Relay output (dry N.O. contact) 
Contact rating: Max. 125 V AC, 2 A  

(Inductive load: COSø = 0.4 or more)

 

Max. 250 V AC, 1 A 

(Inductive load: COSø = 0.4 or more)

Minimum applicable load: 10 mA 

M3.5 terminal Max. 60 V AC/DC: 

  JIS CVV 1.25 mm

2

 or JIS IV 1.25 mm

2

Min. 60 V AC/DC: 
  JIS CVV 2.0 mm

2

 or JIS IV 2.0 mm

2

 

Max. 50 m long 

Totalizer pulse input 

Pulse width:  Min. 30 ms ON time 
 

Min. 30 ms OFF time 

 

Min. 100 ms ON+OFF time 

Applied current/voltage: 5 mA DC typ./12 V DC typ. 
Connectable 

output: Dry contact, transistor output 

 

(open collector output) 

Memory backup: Replaceable lithium battery 

M3.5 terminal JIS CVV 1.25 mm

2

 or JIS IV 1.25 mm

2

 

Max. 50 m long 

1–5 V DC input 

Input impedance: 100 k

 

Isolation: Not isolated from other inputs 

M3.5 terminal JCS CVV-S 1.25 mm

2

 or  

JCS CPEV-S 

⌀0.9 mm 

Max. 50 m long 

4–20 mA DC input 

Input impedance: 300 

 

Isolation: Not isolated from other inputs 

M3.5 terminal JCS CVV-S 1.25 mm

2

 or  

JCS CPEV-S 

⌀0.9 mm 

Max. 50 m long 

Pt100 input 

Input type: Resistance Temperature Detector (Pt100)
Sensing range: -20 °C to 80 °C 

M3.5 terminal JCS CVV-S 1.25 mm

2

  

Max. 50 m long (The difference between 
each wiring must be within 50 cm.) 

Remote control relay drive 
output 

Output type: Thyristor output (dry contact) 
Operating circuit voltage: 24 V AC 
Output rating: Max. 1.5 A 
Number of connection:  1 remote control relay  
  per 

output 

M3.5 terminal JIS CVV 1.25 mm

2

 or JIS IV 1.25 mm

2

 

Max. 50 m long 

Communication 

Transmission method:  Pole response voltage  
  transmission 
Transmission speed: 4800 bps 

Connector*

1

 LAN 

cable*

2

 

Power supply 

Rated voltage: 100 V AC to 240 V AC, 50 Hz/60 Hz 
Operating voltage: 85 V AC to 264 V AC, 50 Hz/60 Hz

M3.5 terminal JIS CVV 2.0 mm

2

 or JIS IV 2.0 mm

2

 

Ground 

Ground resistance: Max. 100 

  

M3.5 terminal JIS CVV 2.0 mm

2

 or JIS IV 2.0 mm

2

 

Notes: 
 *1  For connector connection, use Bel Stewart Connector’s Plug: Model SS-37000-002. 
 

This plug is also available at Azbil Corporation. (Part No. DY7207A0100, 100 pieces/set) 

 *2  LAN cable compliant with EIA/TIA-568 Category 3 or over (

0.5 mm x 4 pairs) is required. 

For *1 and *2, the connector cable (regular cable: Part No. DY7210, short cable: Part No. DY7220) are available at Azbil Corporation. 

Summary of Contents for WY7208W Series

Page 1: ...g with the rotary switch and functional setting with the DIP switch no other setting is required for setup Model Numbers Model number Description WY7208W1120 DIO model Select from the following point...

Page 2: ...ns WARNING Be sure to ground the product with ground resistance of less than 100 Improper grounding might cause electric shock or malfunction Before detaching the terminal cover be sure to turn off th...

Page 3: ...ng so might cause device failure When replacing the battery read the instructions hereinafter and correctly attach the new battery to the product Failure to do so might cause overheating burst or leak...

Page 4: ...dedicated LAN general office purpose LAN CDGP AI Pt TOT model Model WY7208W4140 WY7208W6140 WY7208W3180 Sensor Meter Infilex FC Model WY5205W8000 WY5205W9000 FCU User terminal Model QY7205 QY7290 FCU...

Page 5: ...ime Min 30 ms OFF time Min 100 ms ON OFF time Applied current voltage 5 mA DC typ 12 V DC typ Connectable output Dry contact transistor output open collector output Memory backup Replaceable lithium b...

Page 6: ...AB 5702 6 Dimensions DIO DI AI RRD and Pt models Figure 2 Dimensions DIO DI AI RRD and Pt models TOT model Figure 3 Dimensions TOT model 60 75 95 65 95 65 60 75 19 5...

Page 7: ...grams do not show the appropriate mounting orientation DIO model Point configurations and point assignment to the terminals DIO model Point configuration DO1 1 DO1 2 DI1 1 DI1 2 DO2 1 DO2 2 DI2 1 DI2...

Page 8: ...evice in connection From device in connection From device in connection From device in connection From device in connection Note If the output of the device in connection has polarity connect 0 V wire...

Page 9: ...CN3 SC bus CN4 SC bus From Pt100 temperature sensor From Pt100 temperature sensor From Pt100 temperature sensor To another sub controller Min 20 2 75 x n 35 DIN rail width Min 20 1 2 Min 50 95 Min 50...

Page 10: ...act output x 2 0000 Status alarm point x 4 0000 Status point x 8 0001 Start stop status point maintained contact output x 2 0001 Alarm point x 8 0010 Status point x 4 alarm point x 4 0011 Start stop s...

Page 11: ...r a year or over replace the battery before turning on the CDGP IMPORTANT Lithium battery replacement of TOT model Replace the battery while the CDGP TOT model is being powered Failure to do so will e...

Page 12: ...RD model using the remote control relay and the remote control breaker For providing these devices and the remote control switch thoroughly read the instructions supplied by their manufacturers Do not...

Reviews: